![PerkinElmer Raman Flex 400 Series Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/perkinelmer/raman-flex-400-series/raman-flex-400-series_getting-started-manual_1530229058.webp)
58 . RamanFlex 400 Series Getting Started Guide
Figure 16 System verification/calibration flow diagram
If wavelength calibration verification fails and the Description of the calibration report
spectrum states that peak wavelengths are outside limits, then wavelength and laser
frequency calibrations should be performed. When these calibrations have been completed
successfully, repeat the wavelength verification to ensure that the instrument has been
brought back within specification.
If the wavelength calibration verification fails because the spectral intensity is low, then this
may also be due to the calibration sample holder not being correctly positioned, the sample
not being optimally focused, or because the laser key is not switched on or the laser power is
too low. When you have addressed these issues, repeat the wavelength calibration
verification to ensure that the instrument has been brought back within specification. If the
instrument fails the wavelength calibration verification again, please contact PerkinElmer.
SOP-7: Wavelength Calibration Verification
1.
Ensure the RamanFlex 400 is switched on.
If not, follow SOP-1.
NOTE: If you have a RamanFlex 400F with two fiber optic probes, ensure that the probe not
in use is covered.
2.
Select
Raman Instrument
from the Setup menu.
The Setup Instrument tabs are displayed.
3.
Select the Setup Instrument Advanced tab and ensure that the laser power is set to
100%.
4.
Select
.
The Live tab is displayed.
Wavelength Calibration
Verification
Laser Frequency
Calibration
Wavelength
Calibration
Finished.
Instrument ready
for use
Check Laser Key is
turned on and laser
power is set to
100%
Check position and
focus of polystyrene
sample
Completed
Fail
Peak
wavelengths
outside limits
Fail
Intensity
low
Pass
Содержание Raman Flex 400 Series
Страница 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 24: ...24 RamanFlex 400 Series Getting Started Guide ...
Страница 25: ...Raman History Theory and Instrumentation ...
Страница 32: ...32 RamanFlex 400 Series Getting Started Guide ...
Страница 33: ...Preparing your Facility ...
Страница 37: ...Installing your RamanFlex 400 Series ...
Страница 44: ...44 RamanFlex 400 Series Getting Started Guide ...
Страница 45: ...Using the RamanFlex 400 Series ...
Страница 66: ...66 RamanFlex 400 Series Getting Started Guide ...
Страница 67: ...Maintenance and Troubleshooting ...
Страница 74: ...74 RamanFlex 400 Series Getting Started Guide ...