40 . RamanFlex 400 Series Getting Started Guide
Attaching the External Interlock Override
The RamanFlex 400 Series has an external interlock connection located on the rear panel of
the spectrometer module. To operate the instrument, the external interlock must be
satisfied.
You must connect an external door interlock to the 7-way DIN connector labeled
INTERLOCK on the rear panel of the spectrometer, or attach the External Interlock Override
(part number L1320264) supplied with the instrument.
If your Laser Safety Officer determines that an interlock is required on the door to the room,
you will need to purchase the Raman Laser Safety Kit (part number L1320262). For more
information contact your PerkinElmer representative.
Содержание Raman Flex 400 Series
Страница 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 24: ...24 RamanFlex 400 Series Getting Started Guide ...
Страница 25: ...Raman History Theory and Instrumentation ...
Страница 32: ...32 RamanFlex 400 Series Getting Started Guide ...
Страница 33: ...Preparing your Facility ...
Страница 37: ...Installing your RamanFlex 400 Series ...
Страница 44: ...44 RamanFlex 400 Series Getting Started Guide ...
Страница 45: ...Using the RamanFlex 400 Series ...
Страница 66: ...66 RamanFlex 400 Series Getting Started Guide ...
Страница 67: ...Maintenance and Troubleshooting ...
Страница 74: ...74 RamanFlex 400 Series Getting Started Guide ...