42 . RamanFlex 400 Series Getting Started Guide
Switching on the Instrument
1.
Insert the key into the laser power
keyswitch on the front panel of the spectrometer
module and ensure that it is turned to the ARM position.
The RamanFlex is supplied with two keys. One should be placed in the instrument and
the other should be stored in a safe place. Replacement keys (part number L1323503)
can be obtained from PerkinElmer. A laser safety officer or lab manager can control use
of the instrument with these keys.
2.
If your instrument is fitted with a deep depletion CCD, connect the 7.5 V PSU (part
number L1323525) from the
DD POWER
socket on the rear panel to a mains power
socket.
3.
If your instrument is fitted with a deep depletion CCD, connect the Tygon tubing
between the water recirculator (part number L1323010) and the
WATER IN/OUT
sockets on the rear panel, fill the storage tank of the cooler with water, and then plug in
and switch on the recirculating cooler.
For more information about the recirculating cooler, see the operating manual supplied
with the system.
4.
Connect the instrument power supply unit (part number L1321802) to the spectrometer
Main Power
socket and then switch on.
Do not connect to the Aux Power connector.
Содержание Raman Flex 400 Series
Страница 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 24: ...24 RamanFlex 400 Series Getting Started Guide ...
Страница 25: ...Raman History Theory and Instrumentation ...
Страница 32: ...32 RamanFlex 400 Series Getting Started Guide ...
Страница 33: ...Preparing your Facility ...
Страница 37: ...Installing your RamanFlex 400 Series ...
Страница 44: ...44 RamanFlex 400 Series Getting Started Guide ...
Страница 45: ...Using the RamanFlex 400 Series ...
Страница 66: ...66 RamanFlex 400 Series Getting Started Guide ...
Страница 67: ...Maintenance and Troubleshooting ...
Страница 74: ...74 RamanFlex 400 Series Getting Started Guide ...