52 . RamanFlex 400 Series Getting Started Guide
The Spectrum start-up screen is displayed, followed by a dialog that may require your
login details:
8.
If required, enter your
User name
and
Password
, and then click
OK
.
9.
Select the
Instrument
you want to use and click
Connect
.
OR
If you want to work with data that has been collected previously, without connecting to
an instrument, select
Work Offline
.
Spectrum starts.
If you selected to connect to the instrument, it will prepare for use by turning on the
laser and cooling the detector to its operating temperature (usually −50 °C). Messages
will be displayed in the Status bar at the bottom of the workspace, and in the Scan
icons on the Measurement bar.
NOTE: The CCD detector will become more stable the longer it is cooled.
SOP-2: Turning the Instrument Off
1.
Shut down Spectrum software by selecting Exit from the File
menu to return to the
Windows operating environment.
2.
Switch off the spectrometer by pressing the power switch on the PSU to OFF:
- ON
Ο
- OFF
After Spectrum closes, if you do not switch off the instrument it will remain active for
5 minutes. This is so that you do not have to wait for the CCD detector to cool and the laser
to stabilize should you need to exit Spectrum and then restart the software immediately, or
when changing methods in the AssureID software application.
NOTE: For information on how to use AssureID with a triggered fiber optic probe, refer to
the AssureID on-screen help.
The instrument will shut down automatically after the appropriate time period. Alternatively,
you can shut down the instrument by clicking on the Raman instrument icon in the
Notification area of the Windows Taskbar and accessing the shortcut menu.
You can shut down your PC with the Raman instrument icon displayed. The system will shut
down correctly.
Содержание Raman Flex 400 Series
Страница 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 24: ...24 RamanFlex 400 Series Getting Started Guide ...
Страница 25: ...Raman History Theory and Instrumentation ...
Страница 32: ...32 RamanFlex 400 Series Getting Started Guide ...
Страница 33: ...Preparing your Facility ...
Страница 37: ...Installing your RamanFlex 400 Series ...
Страница 44: ...44 RamanFlex 400 Series Getting Started Guide ...
Страница 45: ...Using the RamanFlex 400 Series ...
Страница 66: ...66 RamanFlex 400 Series Getting Started Guide ...
Страница 67: ...Maintenance and Troubleshooting ...
Страница 74: ...74 RamanFlex 400 Series Getting Started Guide ...