Using the RamanFlex 400 Series . 49
4.
If you have a standard probe, remove the DIN connector in the socket marked LED
FIBER 1 (or LED FIBER 2 if you are removing the second FOP from your RamanFlex
400F).
OR
If you have a triggered probe, disconnect the cable from the USB port on your PC.
5.
Carefully remove the retained ends of the FOP by unscrewing and pulling the fiber optic
cables away from the couplers perpendicularly.
Be extremely careful when removing the FOP as you can easily damage the instrument
and the fiber optic cables.
6.
Replace the plastic caps on the fiber ends of the FOP that have been removed from the
rear panel.
7.
Fasten the retained screw-in caps onto the couplers on the rear panel.
8.
Reconnect the power cables.
Acquiring Spectra with the Fiber Optic Probe
The Raman fiber optic probe has a tight focus spot that is approximately 7.5 mm away from
the lens. The focused spot enables you to pinpoint the measurement area on the sample.
Sampling Solids
Solids can be analyzed by mounting the probe on a stand and adjusting the focal spot on the
surface of the sample. Whenever possible, powders should be compacted down and thin
films of materials should be folded so that the sample covers the entire depth of field. Solids
can be analyzed in glass and plastic containers, similar to liquid samples.
NOTE: Raman spectra are measured in the visible and near-infrared region. Room lights and
sunlight will contaminate your spectra. Take care not to misinterpret room lights or
sunlight as sample features.
Sampling Liquids
Liquid samples can be analyzed by immersing the FOP in the liquid of interest or by analyzing
through the container.
WARNING
The standard FOP is not designed for immersion use. The FOP is
sealed against minor splashes; however, immersion in solvents will
cause serious damage. Ensure an immersion sleeve is in place before
immersing your FOP, see Figure 14. Please contact PerkinElmer if you
have any questions in relation to sampling using the FOP.
Immersion
The FOP can only be immersed in liquid if you have purchased an immersible fiber probe
sleeve. Do not totally submerge the probe. It is only acceptable to immerse the probe up to
1 inch from the top of the sleeve. It may be useful to focus the sample by sliding the FOP
inside the immersion sleeve to obtain the best spectrum. Note that immersion sleeves
contain either quartz or sapphire windows. These windows can yield a detailed Raman
spectrum – do not confuse this with the spectrum of the sample.
Содержание Raman Flex 400 Series
Страница 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 24: ...24 RamanFlex 400 Series Getting Started Guide ...
Страница 25: ...Raman History Theory and Instrumentation ...
Страница 32: ...32 RamanFlex 400 Series Getting Started Guide ...
Страница 33: ...Preparing your Facility ...
Страница 37: ...Installing your RamanFlex 400 Series ...
Страница 44: ...44 RamanFlex 400 Series Getting Started Guide ...
Страница 45: ...Using the RamanFlex 400 Series ...
Страница 66: ...66 RamanFlex 400 Series Getting Started Guide ...
Страница 67: ...Maintenance and Troubleshooting ...
Страница 74: ...74 RamanFlex 400 Series Getting Started Guide ...