Using the RamanFlex 400 Series . 55
Fiber Optic Probe SOPs
NOTE: If you have a RamanFlex 400F with two fiber optic probes, ensure that the probe not
in use is covered.
The SOPs included on the following pages are only appropriate for a standard fiber optic
probe. For details of how to acquire data using a triggered fiber optic probe, see the
Raman
Triggered Fiber Optic Probe
leaflet (L1321940). For details of how to acquire data using a
RamanMicro 300 Accessory, refer to the
Getting Started Guide
supplied with your accessory.
SOP-4: Selecting the Fiber Optic Probe
1.
Ensure the RamanFlex 400 is switched on.
If not, follow SOP-1.
2.
If you did not connect to your instrument on starting Spectrum software, select
Instruments
from the Setup menu and then
Select Instrument
from the sub-menu.
The Instrument Connection dialog is displayed.
3.
Select your Raman instrument and then click
Connect
.
4.
Select
Raman Instrument
from the Setup menu in Spectrum.
The Setup Instrument tabs are displayed in the Dialog pane.
5.
Select the Setup Instrument Basic tab.
6.
Select the Fiber Optic Probe you want to use from the Accessory drop-down list.
The following warning message is displayed.
Содержание Raman Flex 400 Series
Страница 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 24: ...24 RamanFlex 400 Series Getting Started Guide ...
Страница 25: ...Raman History Theory and Instrumentation ...
Страница 32: ...32 RamanFlex 400 Series Getting Started Guide ...
Страница 33: ...Preparing your Facility ...
Страница 37: ...Installing your RamanFlex 400 Series ...
Страница 44: ...44 RamanFlex 400 Series Getting Started Guide ...
Страница 45: ...Using the RamanFlex 400 Series ...
Страница 66: ...66 RamanFlex 400 Series Getting Started Guide ...
Страница 67: ...Maintenance and Troubleshooting ...
Страница 74: ...74 RamanFlex 400 Series Getting Started Guide ...