Clarus 600/680 GC User’s Guide
6-35
Connecting the PPC Modules
The Clarus GC PPC modules have been properly installed and configured at the
factory for your specific combination of injectors and detectors. The configuration
information is stored in a reserved area of the battery-backed memory; however, if
the memory lost the PPC module connect configuration or if you installed a new
PPC module by purchasing an add-on kit, you must follow the procedure(s) listed
in this section.
By connecting a PCC module, you actually set the Clarus GC
controlling software
to recognize a specific PPC module that is installed at a specific location. This is
necessary so that the software can control the installed hardware.
The PPC board has 12 sockets to which various injector and detector modules are
connected. The socket locations and descriptions are listed in the following table
and shown in the following figure:
Table 6-1. Injector and Detector Socket Descriptions
Injector 1
1 (J10)
Injector 2
2 (J11)
Injectors (Carriers)
Detector 1
Air or TCD Ref
3 (J12)
Detector 1
HYD or Make Up
4 (J13)
Detector 1
Detector 2
Air or TCD Ref
5 (J14)
Detector 2
HYD or Make Up
6 (J15)
Detector 2
7 (J16)
8 (J17)
Aux
9 (J18)
10 (J19)
11 (J20)
12 (J21)
Split Vent for CAP & PSS injectors or
Pressure Readout for PKD & POC injectors
Содержание CLARUS 600 GC
Страница 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus 600 680 GC User s Guide ...
Страница 10: ...viii ...
Страница 11: ...Introduction 1 ...
Страница 12: ...Introduction 1 2 ...
Страница 23: ...Touch Screen Navigation 2 ...
Страница 24: ...Touch Screen Navigation 2 2 ...
Страница 32: ...Touch Screen Navigation 2 10 Enabled Disabled When the Stop button is pressed an abbreviated Stop Menu appears ...
Страница 73: ...Using the Active Method 3 ...
Страница 74: ...Using the Active Method 3 2 ...
Страница 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Страница 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Страница 137: ...Setting Up the Detectors 4 ...
Страница 138: ...Setting Up the Detectors 4 2 ...
Страница 192: ...Setting Up the Detectors 4 56 A cross section view of an FPD is shown below Figure 4 31 Flame Photometric Detector FPD ...
Страница 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Страница 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Страница 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Страница 231: ...Using the Method Editor 5 ...
Страница 232: ...Using the Method Editor 5 2 ...
Страница 248: ...Using the Method Editor 5 18 ...
Страница 249: ...Using the Tools Menu 6 ...
Страница 250: ...Using the Tools Menu 6 2 ...
Страница 272: ...Using the Tools Menu 6 24 PID NPD ...
Страница 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Страница 282: ...Using the Tools Menu 6 34 PPC Configure ...
Страница 329: ...Setting up a Typical Analysis 7 ...
Страница 330: ...Setting up a Typical Analysis 7 2 ...
Страница 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Страница 354: ...Setting up a Typical Analysis 7 26 ...
Страница 355: ...Index ...
Страница 356: ...I 2 ...
Страница 361: ......