Using the Active Method
3-8
NOTE:
When using a POC or PSS injector with the subambient accessory, the injector
temperature must not be set below the oven temperature. You can minimize coolant
consumption if using the oven subambient with the PSS or POC installed but not being
used, by deconfiguring the PSS or POC.
Setting a Coolant Cut-In Temperature
The coolant cut-in temperature is the temperature at which the coolant starts to cool the
oven. The default cut-in temperature is 60
°
C. The allowable cut-in temperature range is
30
°
C
-
110
°
C. We recommend that
YOU NOTE THE AMBIENT TEMPERARTURE
AND SET
the cut-in temperature 20 °C above your current ambient temperature.
For example: A room temperature of 20
°
C uses a default temperature of 40
°
C, whereas
a room temperature of 25
°
C uses a 45
°
C cut-in temperature. Examples for setting up
isothermal, 1, and 2 ramp programs follow.
NOTE:
Higher cut-in temperatures cause more coolant to be consumed.
Setting Up an Isothermal Program
Setting up an isothermal program requires entering the following information in the
sequence shown:
•
Initial oven temperature (Temp 1)
•
The total run time for the initial temperature (Time 1)
•
A temperature rate increase of 0 (Rate 1).
Содержание CLARUS 600 GC
Страница 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus 600 680 GC User s Guide ...
Страница 10: ...viii ...
Страница 11: ...Introduction 1 ...
Страница 12: ...Introduction 1 2 ...
Страница 23: ...Touch Screen Navigation 2 ...
Страница 24: ...Touch Screen Navigation 2 2 ...
Страница 32: ...Touch Screen Navigation 2 10 Enabled Disabled When the Stop button is pressed an abbreviated Stop Menu appears ...
Страница 73: ...Using the Active Method 3 ...
Страница 74: ...Using the Active Method 3 2 ...
Страница 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Страница 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Страница 137: ...Setting Up the Detectors 4 ...
Страница 138: ...Setting Up the Detectors 4 2 ...
Страница 192: ...Setting Up the Detectors 4 56 A cross section view of an FPD is shown below Figure 4 31 Flame Photometric Detector FPD ...
Страница 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Страница 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Страница 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Страница 231: ...Using the Method Editor 5 ...
Страница 232: ...Using the Method Editor 5 2 ...
Страница 248: ...Using the Method Editor 5 18 ...
Страница 249: ...Using the Tools Menu 6 ...
Страница 250: ...Using the Tools Menu 6 2 ...
Страница 272: ...Using the Tools Menu 6 24 PID NPD ...
Страница 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Страница 282: ...Using the Tools Menu 6 34 PPC Configure ...
Страница 329: ...Setting up a Typical Analysis 7 ...
Страница 330: ...Setting up a Typical Analysis 7 2 ...
Страница 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Страница 354: ...Setting up a Typical Analysis 7 26 ...
Страница 355: ...Index ...
Страница 356: ...I 2 ...
Страница 361: ......