Clarus 600/680 User’s Guide
4-55
Setting up an FPD
WARNING
Flame Photometric Detectors (FPD) use hydrogen as fuel. If
the hydrogen is turned on without a column attached to the
detector fittings inside the oven, hydrogen could diffuse into the
oven creating the possibility of an explosion. To avoid possible
injury, DO NOT TURN ON THE HYDROGEN unless a column
is attached to the detector fittings inside the oven and all fittings
have been leak tested
.
Before disconnecting a column, make certain that the
HYDROGEN HAS BEEN TURNED OFF.
If two FPDs are installed and only one FPD has a column
connected to it, make certain that you plug the unused detector
inlet fitting with a
1/8-inch stainless steel plug (P/N N930-0061).
NOTE:
To prevent potential damage to the FPD in the manual pneumatics version, the
hydrogen needle valve has been factory-set to a very low setting. Set both the
hydrogen and air inlet pressures to 30 psig, then measure and set the flows as stated
in this chapter.
NOTE:
The PPC pneumatics detector gas modules have been set to OFF at the factory.
FPD Overview
In the Flame Photometric Detector (FPD), the column effluent is mixed with
hydrogen and then burned in air. Light emitted from the flame passes through a
lens, filter, and to a photomultiplier tube, which generates an electrical signal.
The Flame Photometric Detector is used for detecting sulfur, phosphorus, or tin
compounds, which produce chemiluminescent reactions with emissions at
wavelengths characteristic of the S
2
, Sn, and HPO species.
A filter that isolates sulfur emissions (P/N N6000637) is supplied with the detector.
Another filter that isolates phosphorus emissions (P/N N6000981) and a filter that
isolates tin emissions (P/N L4145472) are available from
PerkinElmer
at
www.perkinelmer.com
.
Содержание CLARUS 600 GC
Страница 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus 600 680 GC User s Guide ...
Страница 10: ...viii ...
Страница 11: ...Introduction 1 ...
Страница 12: ...Introduction 1 2 ...
Страница 23: ...Touch Screen Navigation 2 ...
Страница 24: ...Touch Screen Navigation 2 2 ...
Страница 32: ...Touch Screen Navigation 2 10 Enabled Disabled When the Stop button is pressed an abbreviated Stop Menu appears ...
Страница 73: ...Using the Active Method 3 ...
Страница 74: ...Using the Active Method 3 2 ...
Страница 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Страница 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Страница 137: ...Setting Up the Detectors 4 ...
Страница 138: ...Setting Up the Detectors 4 2 ...
Страница 192: ...Setting Up the Detectors 4 56 A cross section view of an FPD is shown below Figure 4 31 Flame Photometric Detector FPD ...
Страница 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Страница 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Страница 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Страница 231: ...Using the Method Editor 5 ...
Страница 232: ...Using the Method Editor 5 2 ...
Страница 248: ...Using the Method Editor 5 18 ...
Страница 249: ...Using the Tools Menu 6 ...
Страница 250: ...Using the Tools Menu 6 2 ...
Страница 272: ...Using the Tools Menu 6 24 PID NPD ...
Страница 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Страница 282: ...Using the Tools Menu 6 34 PPC Configure ...
Страница 329: ...Setting up a Typical Analysis 7 ...
Страница 330: ...Setting up a Typical Analysis 7 2 ...
Страница 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Страница 354: ...Setting up a Typical Analysis 7 26 ...
Страница 355: ...Index ...
Страница 356: ...I 2 ...
Страница 361: ......