Clarus 600/680 User’s Guide
4-31
Equalize the Makeup and Reference Gas Flows
The following procedures describe how to adjust the makeup and reference gas
using PPC pneumatics in the Clarus GC.
CAUTION
The TCD flow module for reference gas is calibrated with
helium at the factory. You must recalibrate the TCD flow
module if you are using a reference gas other than helium.
For optimum performance, apply a minimum flow of 5 mL/min to each of the two
detector channels.
Set the Makeup and Reference Gas
With PPC pneumatics the reference gas is set from the Clarus GC touch screen.
Also, PPC pneumatics provide you with TCD filament protection by monitoring the
carrier gas leak tests or monitoring the gas through the alarm feature and shutting
down the TCD current if a gas leak or no gas at all is detected. The PPC modules are
calibrated at the factory; therefore, to set the flows for an analysis, simply follow
this procedure.
1.
To set the make-up and reference gases touch the
Tools
button and touch
Configuration
from the drop down menu.
Содержание CLARUS 600 GC
Страница 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus 600 680 GC User s Guide ...
Страница 10: ...viii ...
Страница 11: ...Introduction 1 ...
Страница 12: ...Introduction 1 2 ...
Страница 23: ...Touch Screen Navigation 2 ...
Страница 24: ...Touch Screen Navigation 2 2 ...
Страница 32: ...Touch Screen Navigation 2 10 Enabled Disabled When the Stop button is pressed an abbreviated Stop Menu appears ...
Страница 73: ...Using the Active Method 3 ...
Страница 74: ...Using the Active Method 3 2 ...
Страница 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Страница 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Страница 137: ...Setting Up the Detectors 4 ...
Страница 138: ...Setting Up the Detectors 4 2 ...
Страница 192: ...Setting Up the Detectors 4 56 A cross section view of an FPD is shown below Figure 4 31 Flame Photometric Detector FPD ...
Страница 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Страница 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Страница 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Страница 231: ...Using the Method Editor 5 ...
Страница 232: ...Using the Method Editor 5 2 ...
Страница 248: ...Using the Method Editor 5 18 ...
Страница 249: ...Using the Tools Menu 6 ...
Страница 250: ...Using the Tools Menu 6 2 ...
Страница 272: ...Using the Tools Menu 6 24 PID NPD ...
Страница 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Страница 282: ...Using the Tools Menu 6 34 PPC Configure ...
Страница 329: ...Setting up a Typical Analysis 7 ...
Страница 330: ...Setting up a Typical Analysis 7 2 ...
Страница 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Страница 354: ...Setting up a Typical Analysis 7 26 ...
Страница 355: ...Index ...
Страница 356: ...I 2 ...
Страница 361: ......