Clarus 600/680 User’s Guide
4-51
Activating the NPD Bead
The bead requires an activation step prior to use.
Do not inject a sample during
Activation.
To Activate the bead, carefully follow this procedure in sequential steps below:
1.
Turn the bead potentiometer dial clockwise to apply current to the bead.
2.
Slowly increase the NPD control voltage potentiometer to a setting of 700.
Note:
The bead will gradually acquire an orange glow at this point.
3.
Wait for 5 minutes then increase the potentiometer to 750.
4.
After 1 minute move the potentiometer to 800.
5.
After 1 minute move the potentiometer to 850.
6.
After 1 minute move the potentiometer to 900.
7.
After 1 minute reduce the potentiometer to 850.
8.
After 1 minute reduce the potentiometer to 800.
9.
After 1 minute reduce the potentiometer to 750.
10.
After 1 minute reduce the potentiometer down to 700
11.
For Capillary columns:
Wait 10 minutes, go to the NPD detector screen,
then adjust the potentiometer so that the NPD baseline response reads
around 0.5 mV on Range 1. Then set the desired column flow for the
application to be run.
For Packed Columns:
12.
Allow the baseline to stabilize for 30 minutes.
Wait 10 minutes, go to the NPD detector screen,
then adjust the potentiometer down to 500 or below to drop the bead below
the active level. Remove the Detector Cap Nut (P/N N9300061) and
reconnect the column to the detector. Set the desired column flow for the
application to be run, then again increase the bead current potentiometer to
give a reading around 0.5 mV on Range 1.
13.
Using the potentiometer, adjust the baseline between 0.4 and 0.6 mV at
Range 1 and the bead ready to use.
14.
After conditioning the bead remove the Swagelok plug from the detector
fitting and reconnect the column.
Содержание CLARUS 600 GC
Страница 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus 600 680 GC User s Guide ...
Страница 10: ...viii ...
Страница 11: ...Introduction 1 ...
Страница 12: ...Introduction 1 2 ...
Страница 23: ...Touch Screen Navigation 2 ...
Страница 24: ...Touch Screen Navigation 2 2 ...
Страница 32: ...Touch Screen Navigation 2 10 Enabled Disabled When the Stop button is pressed an abbreviated Stop Menu appears ...
Страница 73: ...Using the Active Method 3 ...
Страница 74: ...Using the Active Method 3 2 ...
Страница 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Страница 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Страница 137: ...Setting Up the Detectors 4 ...
Страница 138: ...Setting Up the Detectors 4 2 ...
Страница 192: ...Setting Up the Detectors 4 56 A cross section view of an FPD is shown below Figure 4 31 Flame Photometric Detector FPD ...
Страница 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Страница 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Страница 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Страница 231: ...Using the Method Editor 5 ...
Страница 232: ...Using the Method Editor 5 2 ...
Страница 248: ...Using the Method Editor 5 18 ...
Страница 249: ...Using the Tools Menu 6 ...
Страница 250: ...Using the Tools Menu 6 2 ...
Страница 272: ...Using the Tools Menu 6 24 PID NPD ...
Страница 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Страница 282: ...Using the Tools Menu 6 34 PPC Configure ...
Страница 329: ...Setting up a Typical Analysis 7 ...
Страница 330: ...Setting up a Typical Analysis 7 2 ...
Страница 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Страница 354: ...Setting up a Typical Analysis 7 26 ...
Страница 355: ...Index ...
Страница 356: ...I 2 ...
Страница 361: ......