Clarus 600/680 GC User’s Guide
6-71
The carrier gas flow rate is in units of mL/min (1 to 100mL/min in 0.1mL/min
increments). There is no default as this field will be already filled by a previous
calculation result.
Velocity
This is a selectable field and, if selected, the t0 field is also enabled for input. The
Inlet Press and Flow fields are deselected and are display only.
The carrier gas velocity is in units of cm/s (1 to 500cm/s in 0.1 cm/s increments).
There is no default as this field will be already filled by a previous calculation
result.
Split Flow Calculations Screen
Column flow rate
This field will take its initial value from the Main Screen Flow field. The same
range and units will apply.
If the value is changed in this field, the entry in the Main Screen will also be
updated and the input pressure, velocity and t0 will be recalculated.
Split flow rate
This field will use mL/min units (0 to 500mL/min range in 1ml/min increments).
The default will be 50mL/min.
Column i.d.
This field will take its initial value from the Main Screen Diameter field. The
same range and units will apply.
If the value is changed in this field, the entry in the Main Screen will also be
updated and the active input pressure, flow or velocity and t0 will be
recalculated.
Liner i.d.
This is the internal diameter of the injector liner. The input range will be 0.1 to
6.0mm in 0.1mm increments.
Содержание CLARUS 600 GC
Страница 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus 600 680 GC User s Guide ...
Страница 10: ...viii ...
Страница 11: ...Introduction 1 ...
Страница 12: ...Introduction 1 2 ...
Страница 23: ...Touch Screen Navigation 2 ...
Страница 24: ...Touch Screen Navigation 2 2 ...
Страница 32: ...Touch Screen Navigation 2 10 Enabled Disabled When the Stop button is pressed an abbreviated Stop Menu appears ...
Страница 73: ...Using the Active Method 3 ...
Страница 74: ...Using the Active Method 3 2 ...
Страница 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Страница 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Страница 137: ...Setting Up the Detectors 4 ...
Страница 138: ...Setting Up the Detectors 4 2 ...
Страница 192: ...Setting Up the Detectors 4 56 A cross section view of an FPD is shown below Figure 4 31 Flame Photometric Detector FPD ...
Страница 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Страница 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Страница 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Страница 231: ...Using the Method Editor 5 ...
Страница 232: ...Using the Method Editor 5 2 ...
Страница 248: ...Using the Method Editor 5 18 ...
Страница 249: ...Using the Tools Menu 6 ...
Страница 250: ...Using the Tools Menu 6 2 ...
Страница 272: ...Using the Tools Menu 6 24 PID NPD ...
Страница 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Страница 282: ...Using the Tools Menu 6 34 PPC Configure ...
Страница 329: ...Setting up a Typical Analysis 7 ...
Страница 330: ...Setting up a Typical Analysis 7 2 ...
Страница 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Страница 354: ...Setting up a Typical Analysis 7 26 ...
Страница 355: ...Index ...
Страница 356: ...I 2 ...
Страница 361: ......