Touch Screen Navigation
2-34
After the flame ignites, the FID output level is monitored to confirm ignition
before producing a READY signal. If ignition is not confirmed, the ignition
process repeats up to two more times if needed. If ignition is not achieved after
three attempts, the flame gases are set to off and a dialog box appears displays
the following message:
“
Flame 1 or 2 failed to ignite.”
Touch the
OK
button. The status is made NOT READY and “
FID 1 or 2
flameout”
appears. If auto-ignition is attempted with the FID gases set to zero
flow, the message “
No ignition. Flame gases are off”
appears on the screen. To
proceed, go the PPC screen and touch
Reset
to the Configuration screen and reset
the air and hydrogen to their 450 and 45 mL/min flow rates.
Auto-Ignite
The auto-ignition process only occurs before the start of each run just prior to the
instrument becoming READY. However, the manual flame ignition process,
performed by touching the Ignite button, can be initiated at any time.
Auto-ignite proceeds in the following sequence:
1.
When auto-ignite begins the Clarus GC equilibrates.
2.
Once the two minutes of equilibration expire,
PRE-RUN
begins and the
AutoSystem detects that the flame is not lit and indicates this by alternately
flashing
Flame?
and
Method 5
in the status bar.
3.
After determining that the flame is not lit, the AutoSystem switches to a not
ready state and attempts to automatically ignite the flame.
Flame Ignition 1
appears in the status bar.
4.
The air flow is shut-off and hydrogen is purged through the FID for
10 seconds.
5.
The ignitor energizes, the flame ignites (indicated by a pop sound), and air flow
is returned to sustain the flame. The FID output level is monitored to confirm
that the flame is lit. If a lighted flame is not confirmed, the AutoSystem
repeats the ignition process up to two more times.
Содержание CLARUS 600 GC
Страница 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus 600 680 GC User s Guide ...
Страница 10: ...viii ...
Страница 11: ...Introduction 1 ...
Страница 12: ...Introduction 1 2 ...
Страница 23: ...Touch Screen Navigation 2 ...
Страница 24: ...Touch Screen Navigation 2 2 ...
Страница 32: ...Touch Screen Navigation 2 10 Enabled Disabled When the Stop button is pressed an abbreviated Stop Menu appears ...
Страница 73: ...Using the Active Method 3 ...
Страница 74: ...Using the Active Method 3 2 ...
Страница 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Страница 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Страница 137: ...Setting Up the Detectors 4 ...
Страница 138: ...Setting Up the Detectors 4 2 ...
Страница 192: ...Setting Up the Detectors 4 56 A cross section view of an FPD is shown below Figure 4 31 Flame Photometric Detector FPD ...
Страница 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Страница 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Страница 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Страница 231: ...Using the Method Editor 5 ...
Страница 232: ...Using the Method Editor 5 2 ...
Страница 248: ...Using the Method Editor 5 18 ...
Страница 249: ...Using the Tools Menu 6 ...
Страница 250: ...Using the Tools Menu 6 2 ...
Страница 272: ...Using the Tools Menu 6 24 PID NPD ...
Страница 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Страница 282: ...Using the Tools Menu 6 34 PPC Configure ...
Страница 329: ...Setting up a Typical Analysis 7 ...
Страница 330: ...Setting up a Typical Analysis 7 2 ...
Страница 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Страница 354: ...Setting up a Typical Analysis 7 26 ...
Страница 355: ...Index ...
Страница 356: ...I 2 ...
Страница 361: ......