Clarus 600/680 User’s Guide
4-89
Optimizing a Detector Signal to an Integrator for
Maximum Sample Dynamic Range
The Clarus GC contains improved electronics which significantly reduce amplifier
noise. This allows you to amplify the detector signal without getting a large
increase in the noise level. Therefore, it provides excellent minimal detection (signal
to noise) for all Clarus GC detectors. Amplification is set on the detector page in
the
Atten
field, the attenuation button, and entering an attenuation value from x1
(most signal) to x64 (least signal). The maximum signal produced by the Clarus GC
amplifier is 1 V. To achieve the best sensitivity and widest dynamic range, the
sample with the highest concentration must not exceed the 1-V signal.
When using a flame ionization detector (FID), if you operate at range x1 and
attenuation x64, and the signal saturates (exceeds 1 V), then use range x20 and
determine the appropriate attenuation value (from x64 to x1) that produces a
reasonable size peak. When using a nitrogen phosphorus detector (NPD), the same
will apply. When using an electron capture detector (ECD), the only range setting is
x1. To ensure that the signal does not exceed 1 V, you should run your most
concentrated sample at attenuation x64, then change the attenuation to produce a
reasonable size peak. When using a thermal conductivity detector (TCD) the range
or current setting also affects the output signal. Therefore, start at attenuation x64.
As long as the signal does not exceed 1 V, you can amplify the signal by changing
the attenuation.
Содержание CLARUS 600 GC
Страница 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus 600 680 GC User s Guide ...
Страница 10: ...viii ...
Страница 11: ...Introduction 1 ...
Страница 12: ...Introduction 1 2 ...
Страница 23: ...Touch Screen Navigation 2 ...
Страница 24: ...Touch Screen Navigation 2 2 ...
Страница 32: ...Touch Screen Navigation 2 10 Enabled Disabled When the Stop button is pressed an abbreviated Stop Menu appears ...
Страница 73: ...Using the Active Method 3 ...
Страница 74: ...Using the Active Method 3 2 ...
Страница 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Страница 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Страница 137: ...Setting Up the Detectors 4 ...
Страница 138: ...Setting Up the Detectors 4 2 ...
Страница 192: ...Setting Up the Detectors 4 56 A cross section view of an FPD is shown below Figure 4 31 Flame Photometric Detector FPD ...
Страница 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Страница 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Страница 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Страница 231: ...Using the Method Editor 5 ...
Страница 232: ...Using the Method Editor 5 2 ...
Страница 248: ...Using the Method Editor 5 18 ...
Страница 249: ...Using the Tools Menu 6 ...
Страница 250: ...Using the Tools Menu 6 2 ...
Страница 272: ...Using the Tools Menu 6 24 PID NPD ...
Страница 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Страница 282: ...Using the Tools Menu 6 34 PPC Configure ...
Страница 329: ...Setting up a Typical Analysis 7 ...
Страница 330: ...Setting up a Typical Analysis 7 2 ...
Страница 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Страница 354: ...Setting up a Typical Analysis 7 26 ...
Страница 355: ...Index ...
Страница 356: ...I 2 ...
Страница 361: ......