Clarus 600/680 User’s Guide
3-61
Auto Resume
Check
the
Auto Resume
checkbox to allow the autosampler program to automatically
resume where it left off should a power failure occur.
Running Viscous Samples
The Clarus GC allows you to vary the length of time the plunger pauses at the top-of-
stroke when acquiring samples. In addition, the plunger speed also slows down during
injection. The more viscous the sample, the higher you should set this value. To optimize
reproducibility you should experiment with different values.
The plunger speed during sample acquisition and injection will vary with the viscosity of
the sample. To find the optimum setting for the sample being analyzed, you should
experiment with various syringe settings.
Highlighting the
Viscosity Control
Field use the up and down arrow or keypad
buttons to enter a value between 1 and 15. This value is the number of
seconds the plunger pauses at the top-of-stroke during sample acquisition.
NOTE:
You can obtain improved precision with viscous samples by using the 0.5-mL syringe and
septa in the waste/wash vials in place of the diffusion caps.
Using Solvent Pre-Washes
Pre-washes refers to the number of solvent washes you want the system to perform prior
to the pre-injection sample washes. Highlight the
Solvent Pre-Wash
field and using the
up and down arrow or keypad
buttons type in a value. Acceptable values
are 0, 1 or 2. Pre-washes occur only at the beginning of an autosampler program.
NOTE:
Auto Resume should not be used if you have an FID or FPD configured with PPC, since
the flame will have been extinguished during a power failure.
Priority Injection
Press the
Initiate
button to inject a priority sample during a run.
1.
Touch the Initiate button in the Tasks tab.
2.
In the
Number of Injections
field use the up and down arrow or keypad
buttons to input the number of priority injections. Touch Start.
Содержание CLARUS 600 GC
Страница 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Страница 2: ......
Страница 3: ...Clarus 600 680 GC User s Guide ...
Страница 10: ...viii ...
Страница 11: ...Introduction 1 ...
Страница 12: ...Introduction 1 2 ...
Страница 23: ...Touch Screen Navigation 2 ...
Страница 24: ...Touch Screen Navigation 2 2 ...
Страница 32: ...Touch Screen Navigation 2 10 Enabled Disabled When the Stop button is pressed an abbreviated Stop Menu appears ...
Страница 73: ...Using the Active Method 3 ...
Страница 74: ...Using the Active Method 3 2 ...
Страница 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Страница 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Страница 137: ...Setting Up the Detectors 4 ...
Страница 138: ...Setting Up the Detectors 4 2 ...
Страница 192: ...Setting Up the Detectors 4 56 A cross section view of an FPD is shown below Figure 4 31 Flame Photometric Detector FPD ...
Страница 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Страница 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Страница 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Страница 231: ...Using the Method Editor 5 ...
Страница 232: ...Using the Method Editor 5 2 ...
Страница 248: ...Using the Method Editor 5 18 ...
Страница 249: ...Using the Tools Menu 6 ...
Страница 250: ...Using the Tools Menu 6 2 ...
Страница 272: ...Using the Tools Menu 6 24 PID NPD ...
Страница 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Страница 282: ...Using the Tools Menu 6 34 PPC Configure ...
Страница 329: ...Setting up a Typical Analysis 7 ...
Страница 330: ...Setting up a Typical Analysis 7 2 ...
Страница 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Страница 354: ...Setting up a Typical Analysis 7 26 ...
Страница 355: ...Index ...
Страница 356: ...I 2 ...
Страница 361: ......