DS50 TROUBLE SHOOTING MANUAL LEVEL 3
10. TRANSFERRING EEPROM DATA
8)
Adjust the calibration values: Press the ENTER key to enter the SETUP
menu. Press the UP key until EDIT VARIABLES is displayed, then press
the ENTER key. Use the UP key to select variable number 70, then press
the ENTER key to begin editing. For each variable, use the UP/DOWN
keys to ensure that the number AND the negative sign match the value
from the plate on the crane or from the CD with the calibration values.
Note that holding the button will cause the values to scroll more quickly
than a single button press. After each value matches the desired value,
press the ENTER key to save the value and advance to the next variable.
CONTINUED ON NEXT PAGE
INIT EEPROM
SETUP
EDIT VARIABLES
SETUP
VAR #: 119
EDIT VARIABLES
# 70 VAL:-
20
EDIT VARIABLES