Chapter 6. Maintenance
6
Aurora H2O User’s Manual
6.2.1
Instructions for Optical Absorption Cell Style 1 (cont.)
6.
Use the air blower to blow dry air onto the surface of the mirror. Puff repeatedly until the mirror surface appears
dry (see Figure 90 below).
Figure 90: Drying the Mirror
7.
Repeat steps
4-6
at least three times, using a new lens tissue each time. Each lens tissue wetted with acetone is
used only once to clean off any visible residue on the mirror surface.
8.
Visually inspect the mirror, and record any observation of gross contamination.
9.
If the mirror does not appear to be clean after repeated cleanings, contact BH Technical Support for further
assistance.
10.
If the mirror does appear to be clean, re-install it onto the measurement cell.
Содержание Aurora H2O
Страница 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Страница 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Страница 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Страница 46: ...Aurora H2O User s Manual 22 Figure 33 Aurora H2O Electronic Assembly Wiring Diagram ref dwg 702 8976...
Страница 47: ...Aurora H2O User s Manual 28...
Страница 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Страница 96: ...Aurora H2O User s Manual 36...
Страница 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Страница 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Страница 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Страница 166: ...4 Aurora H2O User s Manual...
Страница 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Страница 169: ......