Chapter 6. Maintenance
2
Aurora H2O User’s Manual
6.2 Cleaning the Mirror (cont.)
2.
Shut down the gas flow through the optical absorption cell.
a.
Turn the inlet isolation needle valve (item 4 in
or
) to the closed position.
b.
Ensure that the sample flow rotatmeter indicates zero flow.
3.
Wear a pair of disposable static dissipative Latex or Nitrile (powder free) gloves.
Note:
There are two versions of the Aurora optical absorption cell assembly (see
Figure 83
below), each with its
own mirror mounting mechanism. The mirror can be removed from the end of this cell for cleaning
WITHOUT DISMOUNTING
the cell. Be sure to follow the correct mirror cleaning procedure for your version of
the optical absorption cell assembly.
Figure 83: Style 1 (Left) and Style 2 (Right) Optical Absorption Cell/Mirror Assemblies
Содержание Aurora H2O
Страница 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Страница 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Страница 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Страница 46: ...Aurora H2O User s Manual 22 Figure 33 Aurora H2O Electronic Assembly Wiring Diagram ref dwg 702 8976...
Страница 47: ...Aurora H2O User s Manual 28...
Страница 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Страница 96: ...Aurora H2O User s Manual 36...
Страница 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Страница 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Страница 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Страница 166: ...4 Aurora H2O User s Manual...
Страница 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Страница 169: ......