Chapter 2. Installation
4
Aurora H2O User’s Manual
2.5 Low Voltage Directive
To comply with the Low Voltage Directive, you must install a switch or circuit breaker on the input power line. For
greatest safety, locate the circuit breaker or power switch near the electronics console.
IMPORTANT:
Installation must be done in accordance with the National Electrical Code, the Canadian Electric Code,
and/or any other applicable local codes.
2.6 Mounting
Use the four mounting tabs to mount the
Aurora H2O
System Assembly in the desired location (see
).
IMPORTANT:
The
Aurora H2O
should only be mounted vertically.
Содержание Aurora H2O
Страница 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Страница 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Страница 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Страница 46: ...Aurora H2O User s Manual 22 Figure 33 Aurora H2O Electronic Assembly Wiring Diagram ref dwg 702 8976...
Страница 47: ...Aurora H2O User s Manual 28...
Страница 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Страница 96: ...Aurora H2O User s Manual 36...
Страница 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Страница 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Страница 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Страница 166: ...4 Aurora H2O User s Manual...
Страница 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Страница 169: ......