Chapter 7. Troubleshooting
7.5 No Flow Measurement Indicated on Aurora H2O Measurement Cell Outlet
Check to make sure that the outlet of the
Aurora H2O
is venting to atmospheric pressure. Ensure that the sample
system valves are configured correctly and that the
Aurora H2O
internal pressure regulator is capable of a barely
positive pressure setting. Check/Replace the filter element in the coalescer/filter as described in Chapter 5,
Maintenance
.
7.6 Verifying Aurora H2O Performance in the Field
There are two methods for verifying the performance of the
Aurora H2O
in the field. The first method is to use a
portable hygrometer such as
BH's PM880
portable hygrometer. The second method is to use a moisture generator or
standard gas cylinder.
7.6.1
Using A Portable Hygrometer
One quick spot check that can be done, that is relatively easy to do and requires minimal set-up time and no
consumables, is to use a second hygrometer.
BH
recommends using a
PM880 Portable Hygrometer
(see
Figure 104
below) for this purpose, with a recently calibrated
Aluminum Oxide Moisture Probe
.
Figure 104: PM880 Portable Hygrometer
The
PM880
can be connected to the outlet of the
Aurora H2O
analyzer, using the portable sample system with the
aluminum oxide moisture probe, to verify the system. The response time of the portable hygrometer is limited, as the
sensor is typically exposed to air during movement to the sample point. So, it is recommended that the sample gas
remains flowing through the portable sample system on the sample outlet of the
Aurora H2O
until equilibrium with
the moisture concentration of the sample gas is reached. You can use the data-logging capability of the PM880 to
determine when steady-state conditions are reached. Consult
BH
if you have application questions regarding this
process.
Содержание Aurora H2O
Страница 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Страница 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Страница 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Страница 46: ...Aurora H2O User s Manual 22 Figure 33 Aurora H2O Electronic Assembly Wiring Diagram ref dwg 702 8976...
Страница 47: ...Aurora H2O User s Manual 28...
Страница 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Страница 96: ...Aurora H2O User s Manual 36...
Страница 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Страница 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Страница 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Страница 166: ...4 Aurora H2O User s Manual...
Страница 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Страница 169: ......