AFX SERIES® OPERATION MANUAL
SECTION 6: FRONT PANEL OPERATION
Entire Contents Copyright
2016 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 106 of 274
6.7.2
USER LIMITS SETTINGS Screen
This screen allows programming user defined voltage and frequency upper and lower limits
to prevent an operator from accidentally programming output settings that could be
damaging to a unit under test.
For example, when testing a 50 Hz transformer, a lower frequency limit setting of 47 would
prevent output frequency programming of values that could cause the transformer to
saturate.
The applies to voltage where a high voltage value could damage a unit under test that was
not designed to handle higher AC or DC input voltage.
Figure 6-17: USER LIMIT SETTINGS Screen
The following parameters can be set from this screen:
Voltage AC
Lower and Upper Vrms set limits for AC programming.
Voltage DC
Lower and Upper Vdc set limits for DC programming.
Frequency
Lower and Upper limits for Frequency programming.
The soft keys on the USER LIMITS SETTINGS screen provide access to additional functionality
as listed in the table below.
SOFT KEY
Description
Apply (SK1)
Accepts new settings and returns to previous screen.
Cancel (SK2)
Returns to the previous screen.
-
-
Back (SK5)
Returns to the previous screen.
Table 6-16: Available USER LIMITS SETTINGS screen soft keys
Содержание 3120AFX
Страница 273: ......