7. CONVERTER INSTALLATION
The maximum transmission length from the sensor is 30 meters. Install the converter within this length.
The converter requires steel pipe 2 inches (50mm) in diameter with furnished U-bolt. See the figure
below.
Fig. 7.1
Locations under any of the following conditions should be avoided:
①
Difficult for inspection and
maintenance.
②
Temperature change and/or
vibration is excessive.
③
Possible immersion in water.
★
Converter must be installed using flameproof
installation procedure.
★
Sensor terminal box must be installed using
intrinsically safe installation procedure.
※
While the stanchion mounting hardware is furnished, the stanchion itself must be provided by the
customer.
Fig. 6.13
〈
Procedure
〉
①
Turn off the power.
②
Remove the cover clamp then remove the
cover. A hex key (JIS B 4648, nominal size 3)
is required. (Fig.6.12)
③
Separate the shield strip. A Phillips screwdriver
is required. (Fig.6.13)
④
Disconnect sensor leads from the terminal block.
A Phillips screwdriver is required. (Fig.6.13)
⑤
Loosen four hex socket head screws securing
the terminal box's neck. A hex key (JIS B
4648, nominal size 4) is required. (Fig.6.13)
⑥
Turn the terminal box to the desired direction,
exercising care not to force the sensor leads.
⑦
After rotating the terminal box, reassemble in
the reverse order of above procedure.
Sensor
Leads
Shield Strip
Cross Recess Pan
Head Screws
Hex Soc.
Head Screw
CAUTION:
Conduit
Connection G3/4
U-bolt
Spring Washer
Hex Bolt
Switching Magnet
Conduit
Connection G3/4