D-160TS-7-E
36
Status output
H/L alarm parameter
F.unit: m
3
/h (standard setting)
Name
Description / Range
Unit
Display
Status output mode
—
R/W
Status output function
—
R/W
Selection item
Description
Normal Open
( )
Status terminals are normally open.
Normal Close
( )
Status terminals are normally closed.
Condition selection of status output.
( ) represents LCD display.
Selection item
Description
Off
(
)
Function disabled.
Error
( )
Status ON during error.
Error+ Alarm
( )
Status ON during error and alarm.
H/L alarm
( )
Status ON during H/L alarm.
Error + H/L
alarm
( )
Status ON during error and H/L alarm.
Function setting of status output.
( ) represents LCD display.
Name
Description / Range
Unit
Display
H/L alarm function
—
R/W
High alarm point
High alarm turns on when instant flow rate is above this setting.
/ 0.0000E+0 to 9.9999E+7
F.unit
R/W
Low alarm point
Low alarm turns on when instant flow rate is below this setting.
/ 0.0000E+0 to 9.9999E+7
F.unit
R/W
H/L alarm hysteresis
Dead-band flow rate setting after an alarm is turned on.
/ 0.0000E+0 to 9.9999E+7
F.unit
R/W
H/L alarm dead time
Alarm dead-time setting. / 0 to 99
sec
R/W
Selection item
Description
Off
( )
Function disabled.
Low alarm
( )
Low alarm only enabled.
High alarm
(
)
High alarm only enabled.
H/L alarm
(
)
Both Low alarm and High alarm
enabled.
Function selection of H/L alarm.
( ) represents LCD display.