PACE 203H – Instructions for Use
5I-17-016X-B-20
63 / 139
Parameter
Range (Step Width)
Unit
ARP
250 (10) 400
ms
Table 49: Range and step width (in pa-
renthesis) for manual setting of ARP
The VRP can be changed with the
softkey labeled
VRP
↑
or the softkey
labeled
VRP
↓
in increments / decre-
ments of 10 ms between 250 ms and
500 ms, respectively, as shown in Table
50.
Parameter
Range (Step Width) Unit
VRP
250 (10) 500
ms
Table 50: Range and step width (in pa-
renthesis) for manual setting of VRP
(Firmware 1.24 or higher)
The parameter changes appear immedi-
ately in the corresponding line of the
lower display (see Table 48) and will
become effective with the next interven-
tion interval.
9.12.5 Setting Options
The Options Menu consists of two parts.
It appears as shown in Table 51 and
Table 52. By pressing the softkey
more...
one can switch between Options Menu 1
and Options Menu 2.
Options Menu 1
Key No.
Beep
1
Alarms
2
Atrial
Trigger
3
more …
4
↑
_
5
Table 51: Options Menu 1
By pressing the softkey labeled
Options
in the Parameters/Options Menu, the
Option Menu 1 for enabling and disa-
bling acoustic signals (beep), alarms and
atrial trigger appears, as shown in Table
51.
The user language can be set via the
Options Menu 2 by pressing the softkey
labeled
Language
as shown in Table 52.
The meaning of the second line in Op-
tions Menu 2 is explained in sections 1.1
(page 78) and 1.1 (page 80).
The statistic feature from the third line is
explained in section 1.1 on page 67.
Содержание PACE 203H
Страница 2: ......
Страница 8: ......
Страница 16: ......
Страница 20: ......
Страница 102: ......
Страница 106: ......
Страница 108: ......
Страница 116: ......
Страница 118: ......
Страница 123: ...PACE 203H Instructions for Use 5I 17 016X B 20 123 139 17 Appendix B Menu Tree Figure 17 Map of Menu Tree Part 1 ...
Страница 124: ...PACE 203H Instructions for Use 124 139 5I 17 016X B 20 Figure 18 Map of Menu Tree Part 2 ...
Страница 130: ......
Страница 132: ......
Страница 136: ......
Страница 138: ......