PACE 203H – Instructions for Use
88 / 139
5I-17-016X-B-20
Abnormal Settings
Reason for
Preventing
Mode
Action
a) Basic Rate
>
MTR
Basic Rate must not
be larger than MTR
DDD
VDD
DAI
VAT
DAT
Transient warning message and
prevent abnormal setting
b) AVD + PVARP + 20ms
>
LRI
To guarantee mini-
mum atrial sensing
phase
DDD
VDD
DAI
VAT
DAT
Transient warning message and
prevent abnormal setting
c) AVD + 60ms
>
LRI
To guarantee mini-
mum VA-delay
D00
DVI
Transient warning message and
prevent abnormal setting
d) PVD + PVARP
≥
URI
(with URI = 1/MTR)
To guarantee
Wenckebach-
behavior
DDD
VDD
DAI
VAT
DAT
Transient warning message and
prevent abnormal setting
e) ARP + 20 ms > LRI
To guarantee mini-
mum atrial sensing
phase
AAI
AAT
Transient warning message and
prevent abnormal settings
Table 60: Abnormal Settings
9.22.7.1 Possible Conflicts
Conflicts can occur in the process of
changing parameter settings. In the tab-
les below, all possible cases are listed
together with the transient conflict warn-
ing message, which will be displayed on
the lower display. In case of a conflict,
the parameter to be changed is limited to
a value still permissible.
Содержание PACE 203H
Страница 2: ......
Страница 8: ......
Страница 16: ......
Страница 20: ......
Страница 102: ......
Страница 106: ......
Страница 108: ......
Страница 116: ......
Страница 118: ......
Страница 123: ...PACE 203H Instructions for Use 5I 17 016X B 20 123 139 17 Appendix B Menu Tree Figure 17 Map of Menu Tree Part 1 ...
Страница 124: ...PACE 203H Instructions for Use 124 139 5I 17 016X B 20 Figure 18 Map of Menu Tree Part 2 ...
Страница 130: ......
Страница 132: ......
Страница 136: ......
Страница 138: ......