The acronyms in the tables have the following definitions:
"
" = Value to be increased
"
" = Value to be decreased
"
A
" = Parameter is set to "Automatic Setting"
"
M
" = Parameter is set to "Manual Setting"
Basic Rate
MTR Message
M
Increase MTR to increase Rate
M
Decrease Rate to decrease MTR
Table 61: Conflicts Type a)
Basic Rate
AVD
PVARP
Message
M
M
Decrease PVARP or AVD to increase Rate
A
M
Decrease PVARP to increase Rate
M
A
Decrease AVD to increase Rate
A
M
Decrease PVARP to decrease Rate
M
A
Decrease AVD to decrease Rate
M
M or A
Decrease PVARP or Rate to increase AVD
M
M or A
Decrease AVD or Rate to increase PVARP
Table 62: Conflicts Type b)
Basic Rate
AVD
Message
M
Decrease AVD to increase Rate
M
Decrease Rate to increase AVD
Table 63: Conflicts Type c)