background image

8

Avoid and Eliminate Condensation

Condensation can cause inaccurate results, and may damage the instrument.  Generally condensation forms 

when the sample, or the instrument itself is brought from a colder location into a moist and warmer one.  All 

must be allowed to come to ambient temperature.  Condensation on tubes must be wiped away until the 

tube sides remain dry.  Never allow condensate or any liquids to enter the sample tube well.  If this should 

happen, take immediate action -- see Maintenance, Section 5.

Always Cap Tubes Before Testing

This seals the contents from contamination, and will prevent varying overhead light from entering the tube 

during measurement.  Wipe sides clean; hold tubes by cap.

Place the 942 on a Horizontal Flat Surface While Testing

Holding it by hand at an angle may disturb the sample and interfere by reflecting the light off the sample’s 

meniscus (the curved surface of the liquid).

Orientation of Sample Tubes in the Tube Well

For reproducible test results when using a 975-TP Plastic Tube, align one of its 2 seams with the “6 o’clock” 

position of the hole.  No. 975-TG Glass Tubes do not have this requirement, and can be inserted without any 

special orientation.

Reagent Tablets

Instructions are given whether the tablet is simply allowed to dissolve, or if it should be crushed with the 

crusher/stirrer rod supplied.  Always allow the tablet-treated test sample to stand for a few seconds so that 

undissolved particles (if any) settle before test reading is taken.

Reagent Stability

For longest shelf life, unless otherwise specifically directed, store reagents in a dry, cool, and dark place.  

Never insert a soiled pipet into a reagent bottle, and make certain to cap bottles tightly.  Write the date of 

receipt on reagent labels, and use previously received reagents first.  Do not use liquid reagents beyond 

their expiration date, if one is provided on the label.  Accurate standard solutions may be used to check the 

reliability of reagents of questionable age.

For Extremely Critical Work

Using a felt-tip pen, place a vertical alignment mark on the upper sides if the tubes (both blank and sample) 

that you intend to use for such tests (but do not extend the lines into the photometric measuring portion of 

the tubes).  When inserting these tubes, rotate them so that the lines align with the white dot at the 6 o’clock 

position of the tube bezel, so they will be identically and reproducibly positioned for each measurement.

Содержание 942-095

Страница 1: ...Series 942 The Mini Analyst Instruction Manual 942 095 Hardness Total Low Range...

Страница 2: ...sing the required reagents by reading the MSDS Material Safety Data Sheets Failure could result in serious injury to the operator or damage to the instrument MSDS www orbeco com CAUTION The accuracy o...

Страница 3: ...General Operating Instructions 10 4 0 Important Reagent Safety and Disposal Precautions 13 4 1 Reagent Handling and Disposal 13 4 2 In Case of Emergency 13 4 3 Explanation of Safety Codes 14 4 4 First...

Страница 4: ...ut 2 line 16 character alphanumeric LCD Zero Set Automatic by microprocessor Standardization Set Automatic by microprocessor Housing Rugged corrosion and water resistant ABS plastic Faceplate Keypad S...

Страница 5: ...d expenses The test procedures in Section 7 are given in step by step easy to follow form with special instructions for your safety and convenience No user adjustments are ever needed LED Narrow Band...

Страница 6: ...er Always replace all 3 batteries at the same time and do not use a mixture of different brands Automatic Power off To conserve battery life the Model 942 shuts itself off automatically 15 minutes aft...

Страница 7: ...event finger marks on the sides of the tubes The containers used to obtain samples and labware used with the instrument should be clean both inside and out To avoid inaccurate results always wash and...

Страница 8: ...this requirement and can be inserted without any special orientation Reagent Tablets Instructions are given whether the tablet is simply allowed to dissolve or if it should be crushed with the crusher...

Страница 9: ...ly the result by the dilution factor 3 For an approximate ranging test measure 5mL or smaller volume into a sample tube and dilute to the 10mL line with distilled or deionized water Perform the test a...

Страница 10: ...afety Data Sheets are provided for all hazardous items Request copies of any you require Protect Model 942 From Temperature Extremes Humidity and Corrosive Gases Do not leave the unit in a vehicle boa...

Страница 11: ...Analyst instruments The sequence of steps covers how to activate the instrument and how to go through its automatic steps to guide you in completing each test After you press the On Off switch the in...

Страница 12: ...ns and Instructions A If display prompts DILUTE RETEST REMOVE SAMPLE Sample is beyond normal range see paragraph 2 3 of section 2 to dilute and then retest again starting with Step 2 above B If displa...

Страница 13: ...instrument will prompt INSERT SAMPLE TUBE meaning the NEXT Sample Tube Insert your next Sample Tube The instrument will now allow you to insert as many Sample Tubes as you wish without having to first...

Страница 14: ...nd or the corresponding Safety Codes comments given below 1 While most of our 68 Mini Analysts use safe reagent tablets it is unavoidable that a few tests use reagents that can be potentially hazardou...

Страница 15: ...skin or eyes copiously flush with water for 15 minutes and if necessary call a physician Clean up and properly dispose of spills immediately Wash hands thoroughly after use 9 Do not store reagents nea...

Страница 16: ...plenty of milk or water immediately Induce vomiting If unconscious or convulsing seek immediate medical attention INHALED Remove exposed person to fresh air at once If difficulty breathing occurs sta...

Страница 17: ...pills should be flushed with plenty of water mopped up and neutralized with 6 Molar Hydrochloric Acid before flushing down the drain with plenty of water Code 3 Chloroform Do NOT flush down drain Stor...

Страница 18: ...everal tests by inserting the blank and continuing as before BATTERIES LOW will continue to appear at the end of each subsequent test until either fresh batteries are installed or the batteries contin...

Страница 19: ...mark Are they positioned all the way down in the tube well D Is the reagent fresh or still valid If questionable retest with fresh reagent E Reviewed any notes in the directions about possible interf...

Страница 20: ...ccident or defects resulting from repairs alterations or installation made by any person or company not authorized by the manufacturer The manufacturer assumes no liability for consequential damage of...

Страница 21: ...er R 3684 If sample is turbid use an analytical quantitative filter paper to filter Samples can be stored if adjusted to pH 2 or less with Nitric Acid Pre Treatment of Sample Tubes This test will dete...

Страница 22: ...t necessarily indicate that hardness is present in this form Notes Note 1 If display reads Dilute and Retest dilute a fresh sample with distilled or deionized water retest and multiply reading by the...

Страница 23: ...23...

Страница 24: ...e Printed in USA 08 08 Rev 1 Orbeco Hellige is a registered trademark of Orbeco Hellige Inc Orbeco Hellige Inc 6456 Parkland Drive Sarasota Fl 34243 Phone 941 756 6410 800 922 5242 Fax 941 727 9654 se...

Отзывы: