5.0   Maintenance and Trouble-Shooting

Each Series 942 Mini Analyst has been designed and constructed to give many years of trouble-free service 

in the lab, plant, classroom, or field.  They are housed in a tough ABS enclosure which is particularly resistant 

to rough handling and chemical attack.  Their key pads are completely sealed against accidental spills.  Treat 

your Mini Analyst as you would any valuable possession.  Do not allow dust, fluids, or condensation to enter 

the tube well, but if this happens, do not open the instrument, instead hold it upside down and gently wipe 

the tube well from below with tissue.  Cover the tube well or the entire instrument when it is not in use to 

prevent debris from entering the well.

5.1   Automatic Self-Diagnosis

Your model 942 automatically performs self-diagnostic steps each time it is turned on and during each 

test -- checking its batteries, analog function, LED Lamp, and Photocell.  Aside from its batteries, it contains 

no user-servicable parts.  If a problem should arise that is not covered by the information  below, contact 

Customer Service by phone at 941.756.6410 or 800.922.5242, by fax at 941.727.9654, or by E-mail at 

[email protected].  

5.2   Replacement of Batteries

The Model 942 will display two warning messages.  The first:  BATTERIES LOW, is a warning that the 

batteries are getting weak.  You may still be able to perform several tests by inserting the blank and

 continuing as before.  BATTERIES LOW will continue to appear at the end of each subsequent test until 

either fresh batteries are installed, or the batteries continue to weaken to the point that the message 

BATTERIES DEAD appears (or the screen remains blank).  In this event no further testing can be performed 

unless the batteries are replaced.

To replace the batteries, turn instrument over, place thumb on mark on battery compartment cover, press 

down and slide the cover off.  Replace all 3 batteries, following polarity markings shown in battery  

compartment.  IMPORTANT:  Rechargeable batteries may be used but cannot be recharged in the instru-

ment.

Do not use a mixture of different battery brands.  Do not mix regular batteries with rechargeable ones.

5.3  Instrument Not Working?

If your Model 942 is not operating at all, or not operating correctly, first review the below points.  (If it

functions correctly, but the results appear wrong, see section 5.4 below).

Check (with a battery checker, not a voltmeter) that each battery has a 1.3v output and that they 

have been installed with the correct polarity as shown in the battery compartment.  Has BLANK 

tube and SAMPLE tube been filled all the way to the 10mL mark?  Are tubes free of obstructions  

such as bubbles or identifying labels?  Outside surfaces of tubes must be unobstructed.

5.4   When Test Results are in Question

If the instrument itself seems to be functioning correctly, but the test results seem to be wrong, first 

thoroughly review the contents of these directions, and the following:

Содержание 942-095

Страница 1: ...Series 942 The Mini Analyst Instruction Manual 942 095 Hardness Total Low Range...

Страница 2: ...sing the required reagents by reading the MSDS Material Safety Data Sheets Failure could result in serious injury to the operator or damage to the instrument MSDS www orbeco com CAUTION The accuracy o...

Страница 3: ...General Operating Instructions 10 4 0 Important Reagent Safety and Disposal Precautions 13 4 1 Reagent Handling and Disposal 13 4 2 In Case of Emergency 13 4 3 Explanation of Safety Codes 14 4 4 First...

Страница 4: ...ut 2 line 16 character alphanumeric LCD Zero Set Automatic by microprocessor Standardization Set Automatic by microprocessor Housing Rugged corrosion and water resistant ABS plastic Faceplate Keypad S...

Страница 5: ...d expenses The test procedures in Section 7 are given in step by step easy to follow form with special instructions for your safety and convenience No user adjustments are ever needed LED Narrow Band...

Страница 6: ...er Always replace all 3 batteries at the same time and do not use a mixture of different brands Automatic Power off To conserve battery life the Model 942 shuts itself off automatically 15 minutes aft...

Страница 7: ...event finger marks on the sides of the tubes The containers used to obtain samples and labware used with the instrument should be clean both inside and out To avoid inaccurate results always wash and...

Страница 8: ...this requirement and can be inserted without any special orientation Reagent Tablets Instructions are given whether the tablet is simply allowed to dissolve or if it should be crushed with the crusher...

Страница 9: ...ly the result by the dilution factor 3 For an approximate ranging test measure 5mL or smaller volume into a sample tube and dilute to the 10mL line with distilled or deionized water Perform the test a...

Страница 10: ...afety Data Sheets are provided for all hazardous items Request copies of any you require Protect Model 942 From Temperature Extremes Humidity and Corrosive Gases Do not leave the unit in a vehicle boa...

Страница 11: ...Analyst instruments The sequence of steps covers how to activate the instrument and how to go through its automatic steps to guide you in completing each test After you press the On Off switch the in...

Страница 12: ...ns and Instructions A If display prompts DILUTE RETEST REMOVE SAMPLE Sample is beyond normal range see paragraph 2 3 of section 2 to dilute and then retest again starting with Step 2 above B If displa...

Страница 13: ...instrument will prompt INSERT SAMPLE TUBE meaning the NEXT Sample Tube Insert your next Sample Tube The instrument will now allow you to insert as many Sample Tubes as you wish without having to first...

Страница 14: ...nd or the corresponding Safety Codes comments given below 1 While most of our 68 Mini Analysts use safe reagent tablets it is unavoidable that a few tests use reagents that can be potentially hazardou...

Страница 15: ...skin or eyes copiously flush with water for 15 minutes and if necessary call a physician Clean up and properly dispose of spills immediately Wash hands thoroughly after use 9 Do not store reagents nea...

Страница 16: ...plenty of milk or water immediately Induce vomiting If unconscious or convulsing seek immediate medical attention INHALED Remove exposed person to fresh air at once If difficulty breathing occurs sta...

Страница 17: ...pills should be flushed with plenty of water mopped up and neutralized with 6 Molar Hydrochloric Acid before flushing down the drain with plenty of water Code 3 Chloroform Do NOT flush down drain Stor...

Страница 18: ...everal tests by inserting the blank and continuing as before BATTERIES LOW will continue to appear at the end of each subsequent test until either fresh batteries are installed or the batteries contin...

Страница 19: ...mark Are they positioned all the way down in the tube well D Is the reagent fresh or still valid If questionable retest with fresh reagent E Reviewed any notes in the directions about possible interf...

Страница 20: ...ccident or defects resulting from repairs alterations or installation made by any person or company not authorized by the manufacturer The manufacturer assumes no liability for consequential damage of...

Страница 21: ...er R 3684 If sample is turbid use an analytical quantitative filter paper to filter Samples can be stored if adjusted to pH 2 or less with Nitric Acid Pre Treatment of Sample Tubes This test will dete...

Страница 22: ...t necessarily indicate that hardness is present in this form Notes Note 1 If display reads Dilute and Retest dilute a fresh sample with distilled or deionized water retest and multiply reading by the...

Страница 23: ...23...

Страница 24: ...e Printed in USA 08 08 Rev 1 Orbeco Hellige is a registered trademark of Orbeco Hellige Inc Orbeco Hellige Inc 6456 Parkland Drive Sarasota Fl 34243 Phone 941 756 6410 800 922 5242 Fax 941 727 9654 se...

Отзывы: