1.0   General Description and Features

Orbeco-Hellige portable Series 942 Specific Test Mini Analyst

 Water Analyzers combine an advanced 

microprocessor-controlled pre-calibrated photometer with all-inclusive “Quick ‘n’ Easy” Reagent Outfits.  You 

can choose from many different complete testing outfits.  They are ideal for every type of water analysis....for  

easy regulatory monitoring, accurate process control, and fast pollution and environmental testing....in the 

lab, plant, classroom, or field.

How They Work.  

A high intensity Light Emitting Diode (LED) transmits a narrow bandwidth light beam 

through the reagent-treated water sample to a light sensitive photovoltiac cell.  Officially approved reagent 

chemistries have been selected that provide a color intensity related to the concentration of the substance 

under test.  A pre-calibrated microprocessor converts the light absorbed directly to numerical test value, 

which is shown on the LCD display.

A Permanent Calibration For Each Mini Analyst

 is stored in the memory of each Model 942 instrument to 

save you time, work and expenses.  The test procedures in Section 7 are given in step-by-step, easy-to-follow 

form, with special instructions for your safety and convenience.  No user adjustments are ever needed.

LED Narrow-Band Light Source

 assures high test sensitivity and very long light source and battery life.

Microprocessor-controlled,

 each 

Mini Analyst

 automatically self-checks its power, circuitry, and lamp readi-

ness at the beginning of each testing session.

The Direct-Reading LCD Display

 provides procedure prompts and exact, unambiguous test results.  It also 

indicated low battery condition.

Model 942 is Powered

 by 3 long-life alkaline AA batteries (supplied).  Battery life is extended by automatic 

shut-off after 15 minutes of non-use.

Accessories Included With the 942

 include all required reagents, 4 scratch-resistant glass tubes, 3 long-life 

AA batteries, and foam-lined carrying case.  Please contact Customer Service for replacement “Quick ‘n’ Easy” 

Reagent Kits, tubes, and other specific test 

Mini Analyst

 available.

Each Model 942 is Made in the U.S.A. and has a 3-year Warranty

For Toll-free Technical Support, 

 contact Orbeco-Hellige, Inc at:

 

Phone:  941.756.6410

                                       800.922.5242 (U.S.A. Only)

               

Fax:        941.727.9654

               

[email protected]

 

www.orbeco.com

              

Specifications

 are listed on the previous page

               

Содержание 942-095

Страница 1: ...Series 942 The Mini Analyst Instruction Manual 942 095 Hardness Total Low Range...

Страница 2: ...sing the required reagents by reading the MSDS Material Safety Data Sheets Failure could result in serious injury to the operator or damage to the instrument MSDS www orbeco com CAUTION The accuracy o...

Страница 3: ...General Operating Instructions 10 4 0 Important Reagent Safety and Disposal Precautions 13 4 1 Reagent Handling and Disposal 13 4 2 In Case of Emergency 13 4 3 Explanation of Safety Codes 14 4 4 First...

Страница 4: ...ut 2 line 16 character alphanumeric LCD Zero Set Automatic by microprocessor Standardization Set Automatic by microprocessor Housing Rugged corrosion and water resistant ABS plastic Faceplate Keypad S...

Страница 5: ...d expenses The test procedures in Section 7 are given in step by step easy to follow form with special instructions for your safety and convenience No user adjustments are ever needed LED Narrow Band...

Страница 6: ...er Always replace all 3 batteries at the same time and do not use a mixture of different brands Automatic Power off To conserve battery life the Model 942 shuts itself off automatically 15 minutes aft...

Страница 7: ...event finger marks on the sides of the tubes The containers used to obtain samples and labware used with the instrument should be clean both inside and out To avoid inaccurate results always wash and...

Страница 8: ...this requirement and can be inserted without any special orientation Reagent Tablets Instructions are given whether the tablet is simply allowed to dissolve or if it should be crushed with the crusher...

Страница 9: ...ly the result by the dilution factor 3 For an approximate ranging test measure 5mL or smaller volume into a sample tube and dilute to the 10mL line with distilled or deionized water Perform the test a...

Страница 10: ...afety Data Sheets are provided for all hazardous items Request copies of any you require Protect Model 942 From Temperature Extremes Humidity and Corrosive Gases Do not leave the unit in a vehicle boa...

Страница 11: ...Analyst instruments The sequence of steps covers how to activate the instrument and how to go through its automatic steps to guide you in completing each test After you press the On Off switch the in...

Страница 12: ...ns and Instructions A If display prompts DILUTE RETEST REMOVE SAMPLE Sample is beyond normal range see paragraph 2 3 of section 2 to dilute and then retest again starting with Step 2 above B If displa...

Страница 13: ...instrument will prompt INSERT SAMPLE TUBE meaning the NEXT Sample Tube Insert your next Sample Tube The instrument will now allow you to insert as many Sample Tubes as you wish without having to first...

Страница 14: ...nd or the corresponding Safety Codes comments given below 1 While most of our 68 Mini Analysts use safe reagent tablets it is unavoidable that a few tests use reagents that can be potentially hazardou...

Страница 15: ...skin or eyes copiously flush with water for 15 minutes and if necessary call a physician Clean up and properly dispose of spills immediately Wash hands thoroughly after use 9 Do not store reagents nea...

Страница 16: ...plenty of milk or water immediately Induce vomiting If unconscious or convulsing seek immediate medical attention INHALED Remove exposed person to fresh air at once If difficulty breathing occurs sta...

Страница 17: ...pills should be flushed with plenty of water mopped up and neutralized with 6 Molar Hydrochloric Acid before flushing down the drain with plenty of water Code 3 Chloroform Do NOT flush down drain Stor...

Страница 18: ...everal tests by inserting the blank and continuing as before BATTERIES LOW will continue to appear at the end of each subsequent test until either fresh batteries are installed or the batteries contin...

Страница 19: ...mark Are they positioned all the way down in the tube well D Is the reagent fresh or still valid If questionable retest with fresh reagent E Reviewed any notes in the directions about possible interf...

Страница 20: ...ccident or defects resulting from repairs alterations or installation made by any person or company not authorized by the manufacturer The manufacturer assumes no liability for consequential damage of...

Страница 21: ...er R 3684 If sample is turbid use an analytical quantitative filter paper to filter Samples can be stored if adjusted to pH 2 or less with Nitric Acid Pre Treatment of Sample Tubes This test will dete...

Страница 22: ...t necessarily indicate that hardness is present in this form Notes Note 1 If display reads Dilute and Retest dilute a fresh sample with distilled or deionized water retest and multiply reading by the...

Страница 23: ...23...

Страница 24: ...e Printed in USA 08 08 Rev 1 Orbeco Hellige is a registered trademark of Orbeco Hellige Inc Orbeco Hellige Inc 6456 Parkland Drive Sarasota Fl 34243 Phone 941 756 6410 800 922 5242 Fax 941 727 9654 se...

Отзывы: