PULSAR Minimal Stress
Installation and operating Manual
OPTIKON 2000
Cod. 111001EN
2012-02-16 Rev.J
8-12
a.
Semi-disposable vitrectome 123001 is reusable and autoclavable approximately 10 times.
b.
Changing a parameter will not undo the default parameters. To recall the default parameters or
to permanently store the new settings in the machine memory, see paragraph "8.7. TO RECALL
OR STORE A PROGRAM".
CAUTION
Vitreous cutter tip should always be test run in a beaker of sterile
solution. Testing in air will cause irreparable damage to the blade.
CAUTION
Aspiration should always be tested in a beaker of sterile solution.
8.5 AUTOMATIC I.V. POLE
If the automatic I.V. pole is connected to the PULSAR, the irrigating solution bottle
will move to the programmed position when the equipment is turned on. The
operator can adjust the bottle position using the UP/DOWN rocker footswitch
and/or the <UP>/<DOWN> soft keys of the Irrigation windows.
To permanently store the new settings in the machine memory, see paragraph "8.7.
TO RECALL OR STORE A PROGRAM".