PULSAR Minimal Stress
Installation and operating Manual
OPTIKON 2000
Cod. 111001EN
2012-02-16 Rev.J
10-2
SYMPTOMS CORRECTIVE
ACTION
Irrigation System
No irrigation flowing:
a) Irrigation pinch valve closed. Check by pressing
the footswitch and verify that the irrigation pinch
valve opens and BSS solution is available.
b) Irrigation tube is disconnected from the
handpiece. Connect irrigation tube to the
handpiece.
c) Kinked or damaged irrigation tubing. Check for
kinks and straighten. Verify damage and replace if
necessary.
d) Closed clamp on connecting tube to irrigation
source. Release clamp to activate irrigation.
e) Vent valve located on drip chamber closed. Open
the vent valve.
Aspiration System
Poor or missing aspiration:
a) I/A tubing set not properly installed. Double check
tubing installation.
b) I/A tubing connectors not properly inserted into
the corresponding connectors on the surgical
handpiece. Insert them properly.
c) I/A tubing set damaged. Exchange for new one.
d) Low "VAC" level. Adjust to slightly higher level.
e) Obstructed tip or handpiece or tubing. Check for
occlusions and clear all debris.
Priming errors
Vac offset:
“VAC OFFSET ERR”
Some vacuum is present although aspiration pump has
not been activated.
a)
Initialization was restarted after a previous error
of the presence of negative pressure. Open the
irrigation clamp and restart priming procedure.
b)
Vacuum sensor needs calibration.
Contact nearest
sales/service office to recalibrate the unit.
Major leakage:
There is a major leakage in the I/A system: