TROUBLESHOOTING GUIDE
10.1 WARNINGS
When the "U/S" Ultrasound function is selected, the following messages may appear
in the status box in the centre of the screen:
ENGLISH ITALIANO
MESSAGE
MESSAGGIO
1
Plug handpiece
Collegare il manipolo
2
Please prime…
Inizializzare il sistema
3
Check tip
Controllare la punta
4
Phaco ready
Ultrasuoni pronti
5 Weak
handpiece
Manipolo poco efficiente
10.2 MISCELLANEOUS PROBLEMS
The troubleshooting guide lists some of the anomalies that may occur in the unit’s
operation, the related symptoms and corrective actions. If the problems with the
PULSAR surgical system persist even after having carried out the corrective actions
below, contact an OPTIKON 2000 authorised service centre.
SYMPTOMS CORRECTIVE
ACTION
Mains power supply
PULSAR completely inoperative:
a) A.C. power cord not connected. Connect power
cord to control console and the mains.
b) A.C. line fuse blown. Replace it.