saline solution to enter the aspiration line. The unit will emit a low audible signal
(800 Hz).
Pressing the main pedal beyond stop 2 activates DIATHERMY, if it has been selected.
The button located above and to the right of the footswitch controls the
CONTINUOUS IRRIGATION; it reverses the condition of the IRRIGATION valve which,
if closed, will open, and if open, will close.
When CONTINUOUS IRRIGATION is activated, gravity causes immediate irrigation to
enter the handpiece.
The audible and visual confirmation signals indicate the modes that have been
selected and the corresponding alarm conditions.
CAUTION
FOOTSWITCH 112102 OPERATES ONLY BY SELECTING THE ITEM “S.
112102” IN THE FOOTSWITCHES AREA.
If you have footswitch 112102, it is advisable to save in the programs
area a work program with footswitch 112102 selected so that it will
not be necessary to reset the footswitch each time the instrument is
turned on.
NOTE:
The Vacuum level is automatically reduced to zero whenever the footswitch is released.
CAUTION
The footswitch is not autoclaveable.
8.7 TO RECALL OR STORE A PROGRAM
All the PULSAR unit's preset parameters can be adjusted independently to meet the
needs of the individual user, who can program the parameters, save up to ten user
programs in the system, and store them for approximately ten years.
The active program's name is indicated in the "ACT. PROG." window located at the
top of the interface screen.
To create and store a new program or modify the active program, first adjust the
unit's various parameters by means of the respective windows within the functions
themselves; an asterisk will appear next to the program's name to indicate that the