PULSAR Minimal Stress
Installation and operating Manual
OPTIKON 2000
Cod. 111001EN
2012-02-16 Rev.J
6-3
2) Gently stretch the supplied irrigation tube and insert
through the perpendicular intake port down into the
side slot of the irrigation pinch valve.
3) Slightly flatten the reflux tube and insert it into reflux
pinch valve. Attach the reflux's female connector to
the "T" connector's side outlet.
4) Line the outlet (tube to the pump) connection of the
Anterior Chamber Surge Suppression System (ACS
3
) to
the dot of "UNLOCK" marking. Engage the ACS
3
in its
receptacle, then rotate it 30° clockwise to line the
outlet to the dot of "LOCK" marking. Confirm that the
cap of the ACS
3
is fully tightened.
5) Insert the tube exiting the ACS3 system into the slot on
the pump's right side, guide the tube around the rotor,
downward, and then through the slot on the pump's
left side. Close the pump's cover.
NOTE:
When using a reusable tubing set, tighten the ACS3 sensor's cover by
screwing it firmly in place.
4)
Close the pump's cover. The tubes' presser will move automatically, locking the
tubes into the proper position. The aspiration line's end, located to the left of
the pump, must be connected to the collection pouch.
5)
Insert the drip chamber spike into the BSS bottle and hand the bottle to the I.V.
pole.
6)
Connect the end of the irrigation and aspiration lines to the corresponding
connectors of the desired surgical handpiece.