When the main pedal is pressed beyond the second mechanical detent (MECH.
DETENT 2), diathermy, if selected, will be activated. If the linear control mode was
selected, a further press on the main pedal will increase the level of diathermy
emission.
"CONTINUOUS IRRIGATION" switch reverses the state of the IRRIGATION pinch
valve which if close opens, if open closes.
The MMP button allows the surgeon to switch between Memory 1, 2, and 3 in U/S
and Memory 1 and 2 in DIATH, I/A, and VIT.
The Change Function button makes it possible to change from U/S to I/A and vice
versa or from VIT to DIATH and vice versa.
When the CONTINUOUS IRRIGATION is activated, there is an immediate
gravity fed irrigation through the handpiece.
Audible and visual feedback is provided as indication of selected modes, and
associated alarm conditions.
The pedal programmed for double linear mode for left-handed users "DBLE. LEFT"
(refer to section 8.7) behaves in the same manner as the double linear mode with
the sole change that the functions set for the horizontal rotation of the pedal have
been inverted.