PULSAR Minimal Stress
Installation and operating Manual
OPTIKON 2000
Cod. 111001EN
2012-02-16 Rev.J
3-4
•
Flammable materials such as disinfecting agents and cleaning agents should be
evaporated before using the bipolar diathermy handpiece. Some materials such
as cotton wool or gauze, if soaked with oxygen, can catch fire because of the
sparks caused by the equipment in its normal use.
•
There is a possible risk for patients carrying cardiac stimulator or stimulating
electrodes because of RF interference caused by diathermy generator. The
stimulator itself can fail. Should any doubt arise, apply to the cardiology
division for advice.
•
Caution: Possible interference with other medical equipment when the bipolar
diathermy handpiece is in use.
Handpieces
•
Do not allow the tip of the vitrectomy probe, diathermy, or I/A handpiece to
touch metallic objects.
•
Do not attempt to alter the shape of any part of a handpiece.
•
Do not attempt to separate the plug from the cable.
•
Do not store or use a handpiece if wet or moist.
I/A
•
Carefully observe the indications of “Irrigation/Aspiration Tubing Set-up”
paragraph in this manual. The failure to follow these indications could have
serious consequences.
•
Correct set-up of irrigation and aspiration lines is critical to ensure proper
operation of the PULSAR unit.
•
Use only original OPTIKON 2000 S.P.A. tubing sets.
•
Do not prime or tune the handpiece while the tip is in the eye of the patient;
this could result in patient injury.
•
Check that the BSS bottle is placed at or above the patient’s eye level.