2-1-5
Timing Charts for each Trigger Mode
There are two methods for Trigger input.
Trigger Input Method
Overview
Trigger Mode
Triggered
Execute Read when the input on
the Parallel TRIG Signal is ON.
•
External Level
•
External Edge
•
Serial Data
Continuous Read
With no Parallel TRIG signal used,
the code reader executes Continu-
ous Read.
•
Continuous Read
•
Continuous Read 1 Output
•
Continuous Read Auto
Below is an Output assignment example and Timing chart.
[Example assignment of OUTPUT signals]
• Output 1: In Read Cycle
It turns ON while the code reader is in its Read cycle.
• Output 2: On Match (or Good Read)
Output Mode: Pulse
It turns ON when there is a Good Read or when it matches with the master symbol (if using the
Matchcode
function).
• Output 3: Mismatch (or on
No Read
)
Output Mode: Pulse
It turns ON when there is a
No Read
or when it does not match with the master symbol (if using
the
Matchcode
function).
For how to set up the Output signal assignments, please refer to
How to Assign the Output Signals
<Timing Chart (External Level)>
1.
Trigger input
(TRIG signal)
2.
Output1
In Read Cycle
3.
Output2
Match
(or Good Read)
4.
Output3
Mismatch
(or
No Read
)
OFF
ON
OFF
ON
OFF
ON
Turns ON on Good Read.
OFF
ON
Read is executed only
while Trigger Input is ON.
Read is executed only
while Trigger Input is ON.
Turns ON
when
No
Read.
<Timing Chart (External Edge)>
1.
Trigger input
(TRIG signal)
2.
Output1
In Read Cycle
3.
Output2
Match
(or Good Read)
4.
Output3
Mismatch
(or
No Read
)
OFF
ON
OFF
ON
OFF
ON
In Read Cycle
Turns ON on Good Read.
Read executed by Trigger Input.
OFF
ON
Turns ON when
No Read
.
2 Controlling Operation and Data Output with Parallel
2 - 7
V430-F Series Autofocus Multicode Reader User Manual for Communication Settings
2-1 Controlling Operation and Data Output
with Parallel
2
2-1-5
T
iming Charts for each T
rigger Mode