2-4 General Information
ChartScan User’s Manual
General
Installation Category:
•
CE:
Category 2 for Line Voltage Input terminal.
All other terminals are Category 1.
Warm
Up:
•
1 hour to rated accuracy.
Expansion Connector:
•
40-conductor connector for connecting expansion chassis via ribbon cable.
ChartScan’s expansion connector (PH401) is located on the analog backplane.
Chassis
Ground
Connection:
•
Screw terminal.
Dimensions:
•
330 mm wide × 241 mm deep × 152 mm high (13” × 9.5” × 6”).
Weight:
•
≈
6.75 kg (15 lbs.) with 4 signal conditioning cards installed
Operating Environment:
•
Standard:
Indoor use, 0 to 50°C;
0 to 95% RH (non-condensing) to
35°C; linearly derate 3% RH/°C
from 35 to 50°C.
•
CE:
Indoor use at altitudes below 2000 m,
5 to 40
°
C; 0 to 80% RH up to 31
°
C
decreasing linearly to 50% RH at 40
°
C.
Control:
•
Power Switch, RS-232 or IEEE 488 option, handshake, parity, baud rate,
calibration memory write enable/disable.
Rear Panel LED Indicators:
•
ALARM, TRIGGER and SCAN; SEND and RECEIVE (for standard serial
interface); ERROR and POWER.
•
TALK, LISTEN and SRQ (on optional IEEE 488 interface card)
Power:
•
105-125V, or 210-250V, 50/60 Hz; 20 VA maximum. Set via internal,
two-position slide switch.
Transient Overvoltage according to Installation Categories:
•
Overvoltage Category II for mains supply. The minimum and normal
category is II.
Pollution Degree:
•
I
−
in accordance with IEC 664
Channels
Number of Option Card Slots:
•
4
Number of Channels:
•
up to 64 channels, expandable to 128 with ChartScan expansion chassis
•
16 channels per card
•
up to 4 cards installed in ChartScan
(up to 4 additional cards with use of ChartScan expansion chassis)
Channel Attributes:
•
High and low set points; hysteresis values for high and low set points.
Scan Sequence:
•
Any combination of temperature and voltage channels may be scanned, but
channels are scanned in ascending numerical order.
Scan Interval:
•
Absolute time between scans (
hh:mm:ss.t
); min = 00:00:00.0,
max = 99:59:59.9.
Note:
Specifying a value of 00:00:00.0 results in no delay
between channel scans.
Содержание OMB-CHARTSCAN 1400
Страница 1: ...August 2002 OMB CHARTSCAN 1400 Portable Data Recorder p n OMB 483 0901 Rev 3 1...
Страница 6: ...iv ChartScan User s Manual...
Страница 24: ...ChartScan User s Manual...
Страница 32: ...2 8 General Information ChartScan User s Manual Notes...
Страница 56: ...3 24 ChartScan User s Manual Notes...
Страница 117: ...ChartScan User s Manual Calibration 6 13 1 Connect an applicable calibration harness to each card Refer to figure below...
Страница 124: ...6 20 Calibration ChartScan User s Manual...
Страница 126: ...A ii ChartScan User s Manual...
Страница 136: ...API Commands Appendix A A 10 ChartScan User s Manual Notes...
Страница 176: ...API Commands Appendix A A 50 ChartScan User s Manual...
Страница 230: ...Appendix C Registers Data Formats Queries ChartScan User s Manual C 13...
Страница 237: ...Registers Data Formats Queries Appendix C C 20 ChartScan User s Manual Notes...
Страница 257: ...E 2 ChartScan User s Manual Notes...
Страница 265: ...ASCII Code Summary Appendix F F 8 ChartScan User s Manual Notes...
Страница 269: ...Abbreviations Appendix H H 2 ChartScan User s Manual Notes...
Страница 271: ...ChartScan User s Manual...