6-2 Calibration
ChartScan User’s Manual
CAUTION
CAUTION
CAUTION
CAUTION
The digital voltmeter (or digital multimeter) used to verify calibration voltage accuracy
must meet the following criteria:
1)
Minimum Resolution: 6-1/2 digits
2)
Minimum DC Accuracy: 0.005% full scale
Failure to comply with this requirement can result in faulty equipment performance and
necessitate additional services of an authorized metrology lab.
CAUTION
CAUTION
CAUTION
CAUTION
Use approved ESD precautions, including static-free work area and grounded wrist strap,
when handling circuit boards and electronic components. Failure to do so could cause
equipment damage due to electrostatic discharge.
The main main unit and each signal conditioning card are calibrated separately.
Note that the main unit and its signal conditioning cards contain components for saving
calibration constants, allowing for card swapping within a system, card swapping between
systems, and system expansion. Calibration relates to the main unit and cards as follows:
Main Unit
: NetScan or ChartScan main chassis, calibrated for offset and gain only.
Signal Conditioning Cards
: Calibrated for offset, gain, and cold junction sensor offset (if thermocouples
are used).
Non-Volatile Storage of Calibration Constants
To preserve calibration constants, main unit constants are stored in an on-board battery backed up RAM
and the constants for each signal conditioning card are stored on the card’s EEPROM.
Hardware Protected RAM
Your scanning device contains a security circuit. This circuit disables writing to the portion of battery
backed up RAM which holds the main unit’s calibration constants and calibration password.
You must
enable the calibration mode prior to calibrating the main unit or changing the password
.
NetScan, Rear Panel
To enable calibration, locate the
Calibration Enable/Disable Pushbutton.
(ChartScan User’s
see notes 1 and 2, below). Push the button so it is in the indent position. This enables
writing to RAM.
Note 1:
ChartScan users only
.
Early production ChartScan units do not have a
Calibration Enable/Disable Pushbutton
.
For units with no calibration pushbutton, a DIP Switch is used instead of the pushbutton, as described in
note 2. ChartScan units that do have the pushbutton will have it located on the rear panel, as in the above
NetScan figure.
Note 2:
ChartScan users only
(for units with no Calibration Enable/Disable Pushbutton)
. If you are using an early
production ChartScan unit, with no calibration pushbutton, then you will enable the calibration mode by
positioning the DIP switch right-most micro-switch (#9) to the “1” (up) position.
Содержание OMB-CHARTSCAN 1400
Страница 1: ...August 2002 OMB CHARTSCAN 1400 Portable Data Recorder p n OMB 483 0901 Rev 3 1...
Страница 6: ...iv ChartScan User s Manual...
Страница 24: ...ChartScan User s Manual...
Страница 32: ...2 8 General Information ChartScan User s Manual Notes...
Страница 56: ...3 24 ChartScan User s Manual Notes...
Страница 117: ...ChartScan User s Manual Calibration 6 13 1 Connect an applicable calibration harness to each card Refer to figure below...
Страница 124: ...6 20 Calibration ChartScan User s Manual...
Страница 126: ...A ii ChartScan User s Manual...
Страница 136: ...API Commands Appendix A A 10 ChartScan User s Manual Notes...
Страница 176: ...API Commands Appendix A A 50 ChartScan User s Manual...
Страница 230: ...Appendix C Registers Data Formats Queries ChartScan User s Manual C 13...
Страница 237: ...Registers Data Formats Queries Appendix C C 20 ChartScan User s Manual Notes...
Страница 257: ...E 2 ChartScan User s Manual Notes...
Страница 265: ...ASCII Code Summary Appendix F F 8 ChartScan User s Manual Notes...
Страница 269: ...Abbreviations Appendix H H 2 ChartScan User s Manual Notes...
Страница 271: ...ChartScan User s Manual...