Appendix C
Registers, Data Formats, & Queries
ChartScan User’s Manual
C-11
U Commands and Descriptions
U0
Causes ChartScan to return a fixed-length status message string. The status message shows the current options in use for
each command.
U1
Causes ChartScan to return a decimal number from 000 to 255. This number is the decimal equivalent of the binary
information on the digital input lines (on the rear panel, card edge digital I/O connector).
U2
Returns information in the Calibration Status Register.
U3
Returns system settings similar to invoking a combination of the following query commands: F?I?L?P?Q?S?T?Y?
U4
Returns high/low/last register data in the current data format for each channel defined by the (C) command, such as the high
reading, the high time stamp, the low reading, the low time stamp, and the last reading.
U5
Returns the same information as U4 but resets the current high/low for all channels.
U6
Returns the following current trigger block information: the trigger block #, current read scan in block, current write scan in
block, trigger time/date stamp, scan # that stop occurred, scan number that end of acquisition occurred, and block completed
flag.
U7
Returns the channels that have been assigned alarm output in the form of: Achan,output where 744 < chan > 1, and 32 <
output > 1.
U8
Returns the channel settings for each configured channel.
U9
Returns the digital input values corresponding to each of the eight digital inputs.
U10
Returns memory size in the form: mmmmm, where m is each digit in the memory size (Kbytes).
U11
Returns the channel number followed by “0” for “not in an alarm state,” or “1” for “in an alarm state. Queried channels are
those having programmed valid alarm set points.
U12
Returns the last date the unit was calibrated in the form: 12:54:00.9,01/28/99
U13
Returns the last scan read.
U14
Returns card ID’s.
U15
Returns the IOtech product information in the form: IOtech, ChartScan, 0, v.r, where v is the version and r is the revision.
U16
Returns measurement parameters.
U17
Returns the RMS value of the last
burst mode
operation.
U18
Returns the contents of the system register.
Status, Event Reporting, and Mask Registers
ChartScan has several internal, eight-bit registers. Each bit represents a certain condition, event or error. The
contents and operation of each of these registers is described in subsequent sections. You can access the
registers via a command issued from the controller.
Status and Event Reporting Registers.
The following four registers cover a wide range of internal error and
event conditions, and each covers a particular aspect of event reporting responsibility.
Calibration Status Register (CSR)
Indicates error conditions during calibration.
Error Source Register (ESC)
Indicates general error conditions with the unit.
Event Status Register (ESR)
Indicates special ‘events’ that have occurred in the unit.
Status Byte Register (STB)
Indicates critical operation events.
Mask Registers.
You can use the following two mask registers to configure the status and event reporting
registers. The mask registers are:
Event Status Enable (ESE)
A mask for the Event Status Register (ESR)
Service Request Enable (SRE)
A mask for the Status Byte Register (STB)
The status registers CSR, ESC and ESR are
Read/Clear-Only
registers (they may only be read and cleared by
the controller). The
read operations
on these registers is a
destructive read
. This is because it clears the
register as it is read. These registers can only be written to by internal ChartScan operations.
The STB status register is a
Read-Only
register (it can only be read by controller). The STB register can only
be cleared or written to by the internal processes of ChartScan.
The mask registers ESE and SRE are
Read/Write/Clear
registers (they can be written and cleared, as well as
read by the controller).
Содержание OMB-CHARTSCAN 1400
Страница 1: ...August 2002 OMB CHARTSCAN 1400 Portable Data Recorder p n OMB 483 0901 Rev 3 1...
Страница 6: ...iv ChartScan User s Manual...
Страница 24: ...ChartScan User s Manual...
Страница 32: ...2 8 General Information ChartScan User s Manual Notes...
Страница 56: ...3 24 ChartScan User s Manual Notes...
Страница 117: ...ChartScan User s Manual Calibration 6 13 1 Connect an applicable calibration harness to each card Refer to figure below...
Страница 124: ...6 20 Calibration ChartScan User s Manual...
Страница 126: ...A ii ChartScan User s Manual...
Страница 136: ...API Commands Appendix A A 10 ChartScan User s Manual Notes...
Страница 176: ...API Commands Appendix A A 50 ChartScan User s Manual...
Страница 230: ...Appendix C Registers Data Formats Queries ChartScan User s Manual C 13...
Страница 237: ...Registers Data Formats Queries Appendix C C 20 ChartScan User s Manual Notes...
Страница 257: ...E 2 ChartScan User s Manual Notes...
Страница 265: ...ASCII Code Summary Appendix F F 8 ChartScan User s Manual Notes...
Страница 269: ...Abbreviations Appendix H H 2 ChartScan User s Manual Notes...
Страница 271: ...ChartScan User s Manual...