Appendix A
API Commands
ChartScan User’s Manual
A-7
Command
Syntax
Description
D#
Set Relay Make Time
D#make
Specify how long to wait after closing a channel relay before
reading valid data, where
make
is the number of 520.833
microsecond intervals.
E
End Calibration Mode
E
Terminate Calibration Mode
E?
Error Status Query
E?
Returns present error condition of the unit with one of the following
error codes:
E000
- No error has occurred.
E001
- Invalid device dependent command (IDDC).
E002
- Invalid device dependent command option (IDDCO).
E004
- Channel configuration error.
E008
- Calibration error.
E016
- Trigger overrun.
E032
- Open TC or Range Error.
E128
- Command conflict error.
F
Set Data Format
Fengr,
format
Configure the input and output formats.
engr
converts the raw data to the engineering units that you
prefer to work with. Valid options are:
0
- degrees Celsius
(default),
1
- degrees Fahrenheit,
2
- degrees Rankine,
3
-
degrees Kelvin, and
4
- Volts.
format
places the readings in a format that you prefer to work
with. Valid options are:
0
- Engineering Units (default),
1
-
Binary (Low Byte/High Byte),
2
- Binary (High Byte/Low Byte),
and
3
- Counts (ASCII).
F?
Query the data format.
F#
Set Burst Mode
Frequency
F#freq
Select burst mode sampling frequency where
freq
is a real
number and where
38.5
<
freq
<
20000.0
Hertz (default).
G
Calibrate Channel Gain
Gchan,
type
Calibrate the correction gain constants for given option card where
chan
is the channel number such that
1
<
chan
<
max
,
where
max = 128
for ChartScan/1400.
The chassis is
chan = 0
.
type
is the channel type.
H
Calibrate Channel
Offset
Hchan
Calibrate the correction offset constants for given option card
where
chan
is the channel number such that
1
<
chan
<
max
, where
max = 128
for ChartScan/1400.
The chassis is
chan = 0
.
I
Set Scan Interval
Inorm, acq
Set the two configurable scan intervals of the unit where
norm
and
acq
are the normal and acquisition scan intervals which
have the form:
hh:mm:ss.t
where
hh
is hours,
mm
is
minutes,
ss
is seconds, and
t
is tenths of a second.
I?
Query scan interval selection.
I#
Digital Input Stamping
I#state
Specify whether or not digital input stamping will be enabled.
state
may have either of the following values: The default of
0
, which disables digital input stamping, or
1
to enable digital
stamping.
J
Calibrate Cold Junction
Offset
Jchan,
type,
temp
Calibrate the cold-junction offsets for cold-junction compensation
of the thermocouple signals where
chan
is the channel number
such that
1
<
chan
<
max
,
where
max = 128
for ChartScan/1400.
type
is the thermocouple channel type.
temp
is the reference temperature currently being applied to the
selected channel, with the form
nnn.n
in degrees
º
C.
K
Enter Calibration Mode
Kkey
Enter Calibration Mode if the keyword (or password) is correct,
where
key
is the keyword in the form of a 5-digit number
nnnnn
. Therefore the password can be no longer than 5
characters.
K?
Query
Kkey
as defined above.
Содержание OMB-CHARTSCAN 1400
Страница 1: ...August 2002 OMB CHARTSCAN 1400 Portable Data Recorder p n OMB 483 0901 Rev 3 1...
Страница 6: ...iv ChartScan User s Manual...
Страница 24: ...ChartScan User s Manual...
Страница 32: ...2 8 General Information ChartScan User s Manual Notes...
Страница 56: ...3 24 ChartScan User s Manual Notes...
Страница 117: ...ChartScan User s Manual Calibration 6 13 1 Connect an applicable calibration harness to each card Refer to figure below...
Страница 124: ...6 20 Calibration ChartScan User s Manual...
Страница 126: ...A ii ChartScan User s Manual...
Страница 136: ...API Commands Appendix A A 10 ChartScan User s Manual Notes...
Страница 176: ...API Commands Appendix A A 50 ChartScan User s Manual...
Страница 230: ...Appendix C Registers Data Formats Queries ChartScan User s Manual C 13...
Страница 237: ...Registers Data Formats Queries Appendix C C 20 ChartScan User s Manual Notes...
Страница 257: ...E 2 ChartScan User s Manual Notes...
Страница 265: ...ASCII Code Summary Appendix F F 8 ChartScan User s Manual Notes...
Страница 269: ...Abbreviations Appendix H H 2 ChartScan User s Manual Notes...
Страница 271: ...ChartScan User s Manual...