6-4 Calibration
ChartScan User’s Manual
Interface Parameters
If your data acquisition device interface is set up for Network (ethernet), you should
change the device interface to RS-232 serial communications prior to performing
calibration. Refer to your hardware setup chapter as needed.
1.a Select the
DIP Switch
toolbar icon to activate the
Select & Verify Interface Setup
window. This
allows you to select and setup the communication interface used for your scanning system.
Select & Verify Interface Setup
1.b Choose the desired interface by clicking the top pull-down triangle (
τ
) and then selecting the applicable
interface from the expanded interface list.
1.c Select the DIP switch image that represents your instrument’s DIP switch overlay.
1.d Click the
Verify
box to make sure the settings are correct.
The
Calibration Enable/Disable pushbutton
(located on the main unit chassis
rear panel
) must be in
the indent (enable position) in order to calibrate the main chassis.
Early production
ChartScan units
(units that have no calibration pushbutton) must have DIP switch
#9 in the 1 (up) position.
After completing the prompt instructions, you can save the calibration settings by selecting the
Save
Interface Parameters
selection (from the
File
pull-down menu or toolbar). The settings and
constants will be stored in a file with a
.CAL
extension.
System Inventory
2.
Select the
Inventory
icon from the toolbar to generate a list of items that make up your scanning
system.
3.
Select the items you wish to calibrate.
Calibration
4.
Select the
Calibration
icon from the toolbar to initiate the automatic calibration process. The
ScanCal
program will prompt you through the calibration. Be ready to supply the corresponding passwords
when prompted by
ScanCal
.
ScanCal
contains a feature that allows you to view the calibration steps without performing an
actual calibration. This is accomplished by using the
Skip this Step
button to skip a procedural
step(s). When using this feature the existing calibration settings will not change.
Содержание OMB-CHARTSCAN 1400
Страница 1: ...August 2002 OMB CHARTSCAN 1400 Portable Data Recorder p n OMB 483 0901 Rev 3 1...
Страница 6: ...iv ChartScan User s Manual...
Страница 24: ...ChartScan User s Manual...
Страница 32: ...2 8 General Information ChartScan User s Manual Notes...
Страница 56: ...3 24 ChartScan User s Manual Notes...
Страница 117: ...ChartScan User s Manual Calibration 6 13 1 Connect an applicable calibration harness to each card Refer to figure below...
Страница 124: ...6 20 Calibration ChartScan User s Manual...
Страница 126: ...A ii ChartScan User s Manual...
Страница 136: ...API Commands Appendix A A 10 ChartScan User s Manual Notes...
Страница 176: ...API Commands Appendix A A 50 ChartScan User s Manual...
Страница 230: ...Appendix C Registers Data Formats Queries ChartScan User s Manual C 13...
Страница 237: ...Registers Data Formats Queries Appendix C C 20 ChartScan User s Manual Notes...
Страница 257: ...E 2 ChartScan User s Manual Notes...
Страница 265: ...ASCII Code Summary Appendix F F 8 ChartScan User s Manual Notes...
Страница 269: ...Abbreviations Appendix H H 2 ChartScan User s Manual Notes...
Страница 271: ...ChartScan User s Manual...