Chapter 5 Operation
81
VISERA VIDEO SYSTEM CENTER OTV-S7V
The color mode setting is retained when the power is turned
OFF and the power cord is disconnected.
5.7
Brightness adjustment
• If the shutter and focus free functions are used while the
brightness is set to maximum and observation takes place for
a prolonged period of time, patient injury may occur.
Always use the minimum level of illumination necessary for
adequate viewing. Whenever possible, avoid close,
stationary viewing and do not leave the distal end of the
endoscope close to mucous membranes for a long time.
Intense endoscopic illumination may cause mucosal burns.
• Do not touch the metal plug of the light guide or the distal end
of the endoscope immediately after use because they will be
extremely hot.
• Follow the instructions below before disconnecting the
camera head or videoscope. Disconnecting the camera head
or videoscope while the brightness is set to maximum could
result in burns or damage the operator’s eyesight.
When using the CLV-S40, set the light source to stand-by
mode or set the AUTO/MANUAL brightness selector to
MANUAL and reduce the light intensity or turn the light
source OFF.
When using the CLD-S, set the AUTO/MANUAL
brightness selector to MANUAL function and reduce the
light intensity. If the light source is turned OFF, it takes a
while to reignite.
When using a light source other than the CLV-S40 or
CLD-S, set the AUTO/MANUAL brightness selector to
MANUAL function and reduce the light intensity or turn
the light source OFF.
Содержание VISERA OTV-S7V
Страница 2: ......
Страница 6: ...Contents iv VISERA VIDEO SYSTEM CENTER OTV S7V...
Страница 10: ...4 Labels and Symbols VISERA VIDEO SYSTEM CENTER OTV S7V...
Страница 18: ...12 Important Information Please Read Before Use VISERA VIDEO SYSTEM CENTER OTV S7V...
Страница 200: ...194 Chapter 7 Troubleshooting VISERA VIDEO SYSTEM CENTER OTV S7V...
Страница 214: ......