Chapter 7 Troubleshooting
187
VISERA VIDEO SYSTEM CENTER OTV-S7V
Chapter 7 Troubleshooting
If the OTV-S7V is visibly damaged, does not function as expected or is found to
have other irregularities during the inspection described in Chapter 3,
“Installation and Connection” and Chapter 4, “Inspection”, do not use the
OTV-S7V. Contact Olympus.
Some problems that appear to be malfunctions may be correctable by referring
to Section 7.1, “Troubleshooting guide”.
If the problem cannot be resolved by the described remedial action, stop using
the instrument and send it to Olympus for repair.
Olympus does not repair accessory parts. If an accessory part becomes
damaged, contact Olympus to purchase a replacement.
Never use the instrument if an abnormality is suspected.
Fatal or serious injury can occur.
7.1
Troubleshooting guide
Irregularity
description
Possible cause
Solution
Power does not
come on.
Power cord not securely plugged into the
OTV-S7V or the wall mains outlet.
Connect power cord
correctly.
A fuse has blown.
Replace both fuses.
Wall mains outlet is not active.
Turn ON the mains
supply for the wall
mains outlet.