Page 19
WIRELESS H2S TRANSMITTER (WT SERIES) - USER GUIDE
9.
SENSOR CALIBRATION
The H2S Transmitter allows users to select the calibration gas concentration to
calibrate the H2S Sensor Module from 10 to 50 ppm in 5 ppm increments. The
transmitter can also calibrate the zero level.
Required Items
- H2S Transmitter
- Bump Test Kit, Part # SX1000-BTK (sold separately)
Housing Adapter
3/16” (ID) Barb
Removable Plug
18” Tubing
- H2S Gas Canister (sold separately),
≤
50 ppm max
1.
Assembly
a.
Properly connect all the components together.
b.
The red plug can be removed or remain connected. Follow your
organization’s best calibration practices.
The sensor must be calibrated on site prior to use due to
local geographical characteristics.
WARNING: Verify that gas concentration in the area has
dissipated to a safe level before performing a sensor
calibration, bump test, or clearing alarms.