APPENDIX B COMMUNICATIONS
7000 Series T72XW Indicators Service ManualB-13
Ohaus Corporation www.ohaus.com
B-16 Commands and Responses MT-SICS Level 1
The following commands of MT-SICS level 1 are available:
SR
Send weight value on weight change (Send and Repeat)
T
Tare
TA
Set or inquiry of preset tare value
TAC
Clear tare value
TI
Tare Immediately
B.16.1 SR
– SEND WEIGHT VALUE ON WEIGHT CHANGE (SEND AND REPEAT)
Command:
SR_Value_Unit
SR
Response:
S_S_ _ _ _ _ 105.1_lb
Current stable weight.
S_D_ _ _ _ _ 106.7_lb
Dynamic weight value.
S_S_ _ _ _ _ 124.3_lb
Next stable weight value.
Response:
S_I
Command understood, not executable at present.
Response:
S_L
Command understood, invalid parameter.
Response:
S_+
Terminal in overload range.
Response:
S_-
Terminal in underload range.
Example
Command:
SR_0.50_kg Send the current stable weight value followed by every load
change > 0.50 kg.
Response:
S_S_ _ _ _ _100.00_kg
Scale stable.
S_D_ _ _ _ _115.23_kg
More than 0.50 kg loaded.
S_S_ _ _ _ _200.00_kg
Scale stable again.
Comments
Command to send the current stable weight once then continuously after every weight change
greater or equal to the “value” a non-stable (dynamic) value followed by the next stable
value. If no preset value is included, the weight change must be at least 12.5% of the last
stable weight value an a minimum of 30d.
SR is overwritten and cancelled by the commands S, SI, SIR, @ and hardware break.