APPENDIX A GEOGRAPHICAL ADJUSTMENT VALUES
Ohaus Corporation www.ohaus.com
B-4
7000 Series T72XW Indicators Service Manual
B.7 Command Formats
Each command received by the terminal via the SICS interface is acknowledged by a response
to the transmitting device. Commands and responses are data strings with a fixed format.
Commands sent to the terminal comprise one or more characters of the ASCII character set.
Commands must be in upper-case.
The parameters of the command must be separated from one another and from the command
name by a space (20xh), in the examples shown in this section, a space is represented as _ ).
Each command must be terminated by <CR>< LF> (0Dxh, 0Axh).
The characters <CR> and <LF>, which can be input using the
ENTER
or RETURN key of most
PC terminal keypads, are not shown in this description; however, it is essential they be included
for communication with the terminal.
Example
Command to tare the terminal:
“TA_20.00_lb” (The command terminators <CR>< LF> are not shown.)
B.8 Response Formats
All responses sent by the terminal to the transmitting device acknowledging the received
commands have one of the following formats:
Response with weight value
Response without weight value
Error message
B.9 Format of the Response with Weight Value
A general description of the response with weight value as follows:
Содержание 7000 Series
Страница 2: ......
Страница 33: ...CHAPTER 1 GETTING STARTED Ohaus Corporation www ohaus com 1 26 7000 Series T72XW Indicator Service Manual...
Страница 65: ...CHAPTER 4 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 4 7000 Series T72XW Indicators Service Manual...
Страница 106: ......
Страница 107: ...30244628 P N 30244628B SERVICE MANUAL 7000 SERIES T72XW INDICATORS...