CHAPTER 2 TROUBLESHOOTING
7000 Series T72XW Indicators Service Manual
2-5
Ohaus Corporation www.ohaus.com
Error Message
Description
power-up zero not
captured
scale in x10 mode
Command not possible while scale is
expanded x10
scale not ready
Print requested and print interlock or auto print
reset requirements have not been met
scale over capacity
Print was requested while scale was over
capacity
scale under zero
Print was requested while scale was blanked
under capacity
tare exceeds limit
Preset tare value is greater than scale capacity
tare not enabled
TARE
key pressed while tare is disabled
tare too small
TARE
key was pressed while weight was less
than 1 display increment
tare under zero
TARE
key was pressed while weight was
below zero
Coin cell voltage low
Contact service
2.4 Troubleshooting
Troubleshooting activities described here are to assist in identifying whether the problem is in
the T72XW terminal or has an external cause.
Blank Display
On/Off Key Test
Coin Cell Battery Test
Problem Diagnosis
Error Codes and Error Messages
Internal Diagnostic Testing
2.4.1 Blank Display
If the display on the T72XW terminal appears to be blank, first confirm if the display is actually
blank or if the backlight is simply off or if the contrast adjustment is incorrect. Even with the
backlight off or a poor contrast setting, changing weight digits should still be barely visible with a
change in weight.