CHAPTER 1
GETTING STARTED
7000 Series T72XW Indicator Service Manual
1-5
Ohaus Corporation www.ohaus.com
1.6
Main Board Wiring Connections
Once the T72XW terminal harsh enclosure is open, connections can be made to the terminal
strips on the main board, as shown in Figure 1-.
Figure 1-2: T72XW Main Board Connections, AC Model
1.6.1
Analog Load Cell Connections
WARNING!
TO AVOID DAMAGE TO THE PCB OR LOAD CELL, REMOVE POWER FROM THE
T72XW TERMINAL AND WAIT AT LEAST 30 SECONDS BEFORE CONNECTING OR
DISCONNECTING ANY HARNESS.
Load cell connections are made to the load cell connector located on the main board as shown
in Figure 1-2.
The AC version of the T72XW terminal is designed to power up to ten 350-ohm load cells (or a
minimum resistance of approximately 35 ohms). The battery version of the T72XW terminal is
designed to power up to four 350-ohm load cells (or a minimum resistance of approximately 87
ohms). To confirm that the load cell load for this installation is within limits, the total scale
resistance (TSR) must be calculated. To calculate TSR:
TSR =
Load Cell Input Resistance (Ohms)
Number of Load Cells
Ensure that the TSR of the load cell network to be connected to the T72XW has a resistance
greater than the minimum values listed above before connecting the load cells. If the resistance
is below the minimum, the T72XW will not operate properly.
DIP
switches
AC
power
connect
or
Load cell
connector
r
Option 1
connector
Option 2
connector