OM0303
Rev 5 July/2021
Page 76 of 108
8.2 Calibration/Service Menu
8.2.1
General
As with the airflow calibration process, the service menu should only be accessed by a Service Technician that is
familiar with the product.
Entry into the Calibration/Service functions requires a service password for entry. After pressing the
Calibrat
ion/Service menu item, a password screen will appear. The default password is “9876”. Once the service
password is entered, the Calibration/Service menu will appear. To exit, either press the menu or home icon.
One additional feature for service technicians is to bypass the whole 2 minute warmup time. By pressing the word
warmup, the system will move directly to run. It is recommended that at least a minute of warmup time be
observed, so the airflows can stabilize at their setpoints to avoid an alarm condition.
Содержание LabGard NU-581-400E
Страница 7: ...OM0303 Rev 5 July 2021 Page 7 of 108 ...
Страница 9: ...OM0303 Rev 5 July 2021 Page 9 of 108 ...
Страница 10: ...OM0303 Rev 5 July 2021 Page 10 of 108 ...
Страница 11: ...OM0303 Rev 5 July 2021 Page 11 of 108 ...
Страница 15: ...OM0303 Rev 5 July 2021 Page 15 of 108 ...
Страница 30: ...OM0303 Rev 5 July 2021 Page 30 of 108 Set New Password Retype new password ...
Страница 43: ...OM0303 Rev 5 July 2021 Page 43 of 108 ...
Страница 101: ...OM0303 Rev 5 July 2021 Page 101 of 108 ...
Страница 102: ...OM0303 Rev 5 July 2021 Page 102 of 108 ...
Страница 103: ...OM0303 Rev 5 July 2021 Page 103 of 108 ...
Страница 104: ...OM0303 Rev 5 July 2021 Page 104 of 108 ...
Страница 105: ...OM0303 Rev 5 July 2021 Page 105 of 108 ...
Страница 106: ...OM0303 Rev 5 July 2021 Page 106 of 108 ...
Страница 107: ...OM0303 Rev 5 July 2021 Page 107 of 108 ...
Страница 108: ...OM0303 Rev 5 July 2021 Page 108 of 108 ...